Table 2. Electroacoustic parameter results and comparison.

Electroacoustic Parameters WHO Recommendation ANSI CTA-2051 LoCHAid Met?
Max OSPL 90 100-130 dB SPL <120 dB SPL 107 dB SPL Yes
OSPL 90 @ 1kHz 90-124 dB SPL NS 90 dB SPL Yes
Average OSPL 90 NS NS 96 dB SPL
Average Gain NS NS 15 dB SPL
Total Harmonic Distortion @ 70dB SPL Input 500 Hz <8%
1000 Hz <8%
1600 Hz <2%
500 Hz <5% 500 Hz = 1%
1000 Hz = 1%
1600 Hz = 1%
Yes
Equivalent Input Noise <30 dB SPL <32 dB SPL 40 dB SPL No
Range of Response and Smoothness 200-8000 Hz
Smoothness—NS
250-5000 Hz
Smoothness
No sharp peaks
<200–>8000 Hz
Smooth
Yes
Battery Life 2-3 Weeks NS 20 days
(with 2 AA batteries)
Yes
HFA (1, 1.5, 3 kHz) @ OSPL 90 NS NS 93 dB SPL
HFA (4,5,6 kHz) @ OSPL 90 NS NS 103 dB SPL

The table lists the ANSI Parameters (OSPL 90, OSPL 60, Total Harmonic Distortion, High Frequency Average, Average Gain, Max OSPL 90) that were tested on the LoCHAid, the WHO Recommendations Preferred Product Profile (PPP) for the device, the ANSI/CTA-2051 recommendations, and the results from the LoCHAID, and whether the targets were met or not for both sets of recommendations [12, 27]. The device is able to meet all the targets except for Equivalent Input Noise. See Discussion in main text about EIN. *NS = Not Specified.