Skip to main content
. 2020 Oct 1;40(5):373-381. doi: 10.5144/0256-4947.2020.373

Table 5.

Limit of SARS-COV-2 detection after automated RNA extraction using the modified DNA kit.

TaqPathTM Assay In-house Assay
Titration (copies) Target Cq Titration (copies) Target Cq
15000 N 23.1 3000 E 31.08
S 29.64 RdRp2 29.35
ORF1b 22.58 RdRp4 29.99
7500 N 22.35 1500 E 32.09
S 32.36 RdRp2 31.53
ORF1b 23.19 RdRp4 29.98
3750 N 22.46 750 E 31.04
S 30.73 RdRp2 29.58
ORF1b 23.75 RdRp4 30.31
1875 N 25.44 375 E 31.63
S 30.35 RdRp2 30.51
ORF1b 24.76 RdRp4 31.43
937.5 N 26.73 187 E 35.61
S 0.00 RdRp2 32.96
ORF1b 25.22 RdRp4 34.82
468.75 N 27.38 93 E 34.86
S 34.56 RdRp2 34.70
ORF1b 23.58 RdRp4 35.35
234.37 N 28.45 46 E 36.55
S 39.92 RdRp2 34.86
ORF1b 28.07 RdRp4 34.97
117.18 N 29.39 23 E 38.23
S 0.00 RdRp2 36.21
ORF1b 29.39 RdRp4 35.84
58.59 N Gene 29.58 11 E 37.30
S Gene 36.11 RdRp2 35.35
ORF 1b 29.40 RdRp4 38.17