Skip to main content
. 2020 Sep 22;58(10):e01669-20. doi: 10.1128/JCM.01669-20

TABLE 2.

Agreement analysis between the Aptima SARS-CoV-2 TMA assay and the Panther Fusion SARS-CoV-2 RT-PCR assay for nasopharyngeal swab specimensa

Aptima SARS-CoV-2 TMA result Panther Fusion SARS-CoV-2 RT-PCR assay result
Positive Negative Total
Positive 64 1 65
Negative 0 75 75
Total 64 76 140
a

Overall agreement was 99.3% (95% CI, 96.1% to 99.9%), positive agreement was 100% (95% CI, 94.3% to 100%), and negative agreement was 98.7% (95% CI, 92.9% to 99.8%).