TABLE 3.
SARS-CoV-2 control vendor | Panther Fusion SARS-CoV-2 RT-PCR result |
||
---|---|---|---|
Concn (c/ml) | Aptima SARS-CoV-2/TMA (n/N) (%) | Panther Fusion SARS-CoV-2 RT-PCR (n/N) (%) | |
Exact Diagnostics SARS-CoV-2 standard | 833 | 40/40 (100) | 38/38 (100) |
417 | 40/40 (100) | 39/39 (100) | |
194 | 40/40 (100) | 40/40 (100) | |
83 | 40/40 (100) | 40/40 (100) | |
19 | 35/39 (90) | 30/40 (75) | |
8 | 29/39 (74) | 21/40 (53) | |
SeraCare Accuplex SARS-CoV-2 verification panel | 833 | 20/20 (100) | 20/20 (100) |
417 | 19/19 (100) | 20/20 (100) | |
194 | 19/19 (100) | 20/20 (100) | |
83 | 20/20 (100) | 18/20 (90) | |
19 | 16/40 (40) | 11/40 (28) | |
8 | 7/40 (18) | 4/40 (10) | |
ZeptoMetrics SARS-CoV-2 (recombinant) | 833 | 39/39 (100) | 40/40 (100) |
417 | 40/40 (100) | 40/40 (100) | |
194 | 37/37 (100) | 40/40 (100) | |
83 | 39/40 (97.5) | 32/40 (80) | |
19 | 12/40 (30) | 14/40 (35) | |
8 | 12/40 (30) | 8/40 (20) |
The lowest concentration of each panel with 100% positivity is shown in bold type. n/N, number positive/number tested.