TABLE 3.

Performance of Aptima SARS-CoV-2 TMA and Panther Fusion SARS-CoV-2 RT-PCR assays for detection of commercially available SARS-CoV-2 controlsa

SARS-CoV-2 control vendor Panther Fusion SARS-CoV-2 RT-PCR result
Concn (c/ml) Aptima SARS-CoV-2/TMA (n/N) (%) Panther Fusion SARS-CoV-2 RT-PCR (n/N) (%)
Exact Diagnostics SARS-CoV-2 standard 833 40/40 (100) 38/38 (100)
417 40/40 (100) 39/39 (100)
194 40/40 (100) 40/40 (100)
83 40/40 (100) 40/40 (100)
19 35/39 (90) 30/40 (75)
8 29/39 (74) 21/40 (53)
SeraCare Accuplex SARS-CoV-2 verification panel 833 20/20 (100) 20/20 (100)
417 19/19 (100) 20/20 (100)
194 19/19 (100) 20/20 (100)
83 20/20 (100) 18/20 (90)
19 16/40 (40) 11/40 (28)
8 7/40 (18) 4/40 (10)
ZeptoMetrics SARS-CoV-2 (recombinant) 833 39/39 (100) 40/40 (100)
417 40/40 (100) 40/40 (100)
194 37/37 (100) 40/40 (100)
83 39/40 (97.5) 32/40 (80)
19 12/40 (30) 14/40 (35)
8 12/40 (30) 8/40 (20)
a

The lowest concentration of each panel with 100% positivity is shown in bold type. n/N, number positive/number tested.