Skip to main content
. 2020 Sep 22;58(10):e01033-20. doi: 10.1128/JCM.01033-20

TABLE 4.

Performance of the Novodiag Bacterial GE+

Pathogen (no. of samples) Novodiag Bacterial GE+ result No. of CRS tests with indicated resulta
Positive Negative Total
Campylobacter sp. (251) Positive 94 0 94
Negative 1 156 157
Salmonella sp. (251) Positive 46 0 46
Negative 0 195 195
EHEC (123) Positive 28 0 28
Negative 0 95 95
ETEC (119) Positive 15 2 17
Negative 0 102 102
Y. enterocolitica (251) Positive 8 0 8
Negative 0 243 243
Shigella sp./EIEC (251) Positive 9 0 9
Negative 0 242 242
a

The CRS was defined as positive when culture was positive or, in the case of a negative culture (or if conventional culture was not performed or unable to detect the pathogen [i.e., EHEC, EPEC, EAEC, ETEC, C. difficile]), when the result was concordant with the one previously obtained with the BD Max extended enteric bacterial panel. When a discrepancy was noticed, CRS was defined as positive when at least one independent PCR assay used for confirmation was concordant with Novodiag Bacterial GE+.