Table 1.

Results of the SD-Biosensor antigen (R-Ag) test compared to RT-PCR in the samples tested. R-Ag sensitivity, specificity, negative and positive predictive values were 70.6 %, 100 %, 87.4 % and 100 %, respectively, compared with RT-PCR. Concordance between the two techniques was.90.3 %.

R-Ag +
N.
R-Ag –
N.
Total
N.
Positive
%
Negative
%
RT-PCR+ 77 32 109 70.6
RT-PCR- 0 221 221 100
Total 77 253 330 23.3