Table 2.

Acquisition and reconstruction parameters used in the phantom experiments.

Scan parameters SECT DECT PCD-CT
PCD-CT
Four-energy mode Two-energy mode
kV 120 80/Sn140 100/Sn140 140 140
Detector type EID EID PCD PCD
Collimation (mm) 128×0.6 32×0.6 32×0.5 32×0.5
Pitch 0.7 0.6 0.6 0.6
Thresholds (keV) N/A N/A [20,31,55,90] [25, 65]
[20,25,57,77]
[20,39,64,89]
[25,45,65,85]
Image thickness/interval (mm) 1.5/1.0 1.5/1.0 1.5/1.0 1.5/1.0
Recon kernel B40 D30 D30 D30
Recon FOV (mm) 200 200 200 200
CTDIvol (mGy) RD: 3.82 RD: 3.52 RD: 3.53 RD: 8.68 RD: 3.88
HD: 18.50 HD: 28.13 HD: 60.74 HD: 60.74

Note: EID, energy integrating detector; PCD, photon-counting detector; FOV, field of view; RD, routine radiation dose; HD, highest radiation dose. Dose values are in terms of volume CT dose index (CTDIvol) and are based on the 32-cm diameter phantom.