Table 1. Final LC–MS Parameters Used for the Last Method Development and Validation Stepsa.

  UPLC-MRM/MS
LC eluent A: H2 0 5% ACN 0.1% FA
  B: ACN 0.1% FA
analytical separation BEH C18, 2.1 × 50 mm, 1.7 μm, 0.5 mL/min, 5% B, 40 °C on column, 5% B for 0.1 min, 5 to 60% B in 1.9 min, 60 to 98% B in 0.5 min, 98 to 5% B in 0.5 min and 5% B for 0.5 min
ESI parameters ESI positive mode
  capillary voltage: +3 kV, cone voltage: 48 V
  source at 150 °C, desolvation at 250 °C
  cone gas flow at 150 L/h, desolvation gas at 1000 L/h, collision gas flow at 0.15 mL/min and nebulizer gas at 7 bar
acquisition parameters MRM mode transitions [m/z] collision energy
  LC1 896.06 > 807.55 24 V
    896.06 > 982.26 24 V
  SIL-LC1 899.60 > 814.44 24 V
    899.60 > 989.27 24 V
data analyses Targetlynx v. 4.1
  regression model: linear with 1/x weighting and origin exclusion. Savitzky–Golay smoothing (iterations: 2, width: 2)
  integration using apex track algorithm
a

ACN: acetonitrile; ESI: Electrospray ionization; n.a. not applicable.