Characteristics of the XAC1051-2qPCR assay performed using three different devices

Plant matrix StepOnePlus QS LC480
qPCR Efficiency lemon 93% 94% 97%
orange 102% 103% 106%
R2 lemon 0.992 0.981 0.984
orange 0.988 0.986 0.987
Intra-assay variation Cv range (median)

0.07–2.48%

(0.34)

0.13–4.6%

(0.60)

0.14–3.6%

(0.81)

Inter-assay variation Cv range (median)

0.49–2.5%

(1.50)

0.55–3%

(1.30)

0.84–2.4%

(1.50)

Cut-off values 35.9 38.43 38.89

LOD95%

(CI)

2.90

(2.76–3.04)

3.04

(2.96–3.11)

3.09

(2.92–3.25)

Vic Ct range 25.8–31.2 22.3–27.9 23.4–29.7