Table 2. Accuracy indices for the rapid test BIOSYNEX COVID-19 assay for SARS-CoV-2 IgM and IgG detection compared to the MAGLUMI 2019-nCoV CLIA.

Biosynex Covid-19 BSS
vs
MAGLUMI 2019-nCov
SARS-CoV-2 IgM SARS-CoV-2 IgG
Sensitivity (95%CI), % 83 (0.64-0.94) 82 (0.67-0.92)
Specificity (95%CI), % 88 (0.74-0.96) 93 (0.75-0.99)
PPV (95%CI) % 83 (0.64-0.94) 95 (0.82-0.99)
NPV (95%CI), % 88 (0.74-0.96) 75 (0.57-0.88)
LR+ (95%CI) 6.79 (2.93-15.69) 10.63 (2.79-40.57)
C-k (95%CI) 0.71 (0.54-0.87) 0.71 (0.54-0.87)
PPV - Positive predictive value. NPV - Negative predictive value. LR - Likelihood ratio. CI - Confidence interval. C-k - Cohen’s kappa.