Table 5.

Result of paired observation of VHW-client interaction

Variable Period VHWs competent with using the device
%, n = 118 95% CI McNemar’s p value
Properly positons the arm of the client before applying the device Baseline 9 5–13 < 0.001
End line 95 90–99
Properly applies the device’s cuff to the client’s arm Baseline 38 30–46 < 0.001
End line 95 90–98
Hears the beep sound of the device before pumping the inflation bulb Baseline 76 67–84 < 0.001
End line 99 98–100
Stops inflating the bulb after hearing the second beep sound Baseline 75 67–83 < 0.001
End line 99 98–100
Can identify the BP readings on the device Baseline 94 90–98 0.04
End line 99 98–100
Understands traffic color on the device and action to take Baseline 84 78–90 0.04
End line 93 89–97
Can define error messages on the device Baseline 41 33–49 < 0.001
End line 82 75–88