Table 1:
Entrance skin dose and mean glandular dose for tungsten target and rhodium and silver filter combination materials and various kV settings for the Hologic digital mammography imaging system.
Target/Filter – kV | W/Rh | W/Ag | ||||
---|---|---|---|---|---|---|
mAs | ESD (mGy) | MGD (mGy) | mAs | ESD (mGy) | MGD (mGy) | |
28 | 118 | 5.01 | 1.2 | 69.6 | 3.5 | 0.72 |
30 | 89.8 | 4.6 | 1.2 | 52.7 | 3.2 | 0.65 |
32 | 67.7 | 4.02 | 1.1 | 40.8 | 2.9 | 0.61 |
34 | 49.3 | 3.3 | 0.94 | 32.2 | 2.6 | 0.56 |
ESD: Entrance skin dose, MGD: Mean glandular dose, W: Tungsten target, Rh: Rhodium, Ag: Silver