Skip to main content
. 2020 Aug 14;36(4):399–408. doi: 10.1007/s12550-020-00405-9

Table 1.

Influence of microfiltration on the chemical composition of a complex mixture of Alternaria toxins, as obtained by LC-MS/MS analysis. Concentrations are provided in micrograms per liter or milligrams per liter except for STTX-III, where absolute quantification was not possible due to a lack of a reference standard; thus, the respective peak area units (AU) are presented

AOH (μg/L) AME (μg/L) ALT (μg/L) TeA (mg/L) TEN (μg/L) ATX-I (mg/L) ATX-II (mg/L) ALP (mg/L) STTX-III (106 AU) ALS (μg/L) AST (mg/L)
Unfiltered 90.3 88.3 89.2 62.5 1.5 1.04 1.38 1.38 680 25.6 2.23
PTFE 84.4 72.7 83.3 60.1 0.9 1.05 1.35 1.30 700 23.3 1.37
PVDF 68.9 < 0.06 75.8 58.0 0.8 0.79 0.69 1.03 254 25.8 < 0.0002
Nylon < 0.3 < 0.06 26.0 56.3 0.8 < 0.00024 < 0.001 < 0.001 5 8.9 < 0.0002
PES < 0.3 < 0.06 15.7 59.3 0.9 < 0.00024 < 0.001 < 0.001 3 7.4 < 0.0002
GF/CA < 0.3 < 0.06 52.7 57.9 0.6 0.0035 < 0.001 0.006 9 23.6 1.10
RC 42.9 38.8 91.7 59.1 0.7 0.98 1.34 1.32 688 23.3 0.71