Performance characteristics of the VISITECT® CD4 Test.

Flow cytometry
VISITECT®CD4 test (visual reading)

Venous blood

Finger-prick blood
≤350 cells/µL >350 cells/µL ≤350 cells/µL >350 cells/µL
≤350 cells/µL (145) 140 5 123 22
>350 cells/µL (55) 16* 39* 15 40
Sensitivity, % (95%CI) 96.6% (92.1% to 98.9%) 84.8% (77.9% to 90.2%)
Specificity, % (95%CI) 70.9% (57.1% to 82.4%) 72.7% (59.0% to 83.9%)
PPV, % (95%CI) 89.7% (83.9% to 94.0%) 89.1% (82.7% to 93.8%)
NPV, % (95%CI) 88.6% (75.4% to 96.2%) 64.5% (51.3% to 76.3%)

*The results of the second (duplicate) VISITECT®CD4 test using venous blood were 17 and 38 participants with ≤350 and >350 cells/μL, respectively, giving a sensitivity of 96.6% (92.1% to 98.9%), a specificity of 69.1% (55.2% to 80.9%), a PPV of 89.2% (83.2% to 93.6%) and a NPV of 88.4% (74.9% to 96.1%).