Table 2.
Measurements of volume, plateau pressure and PEEP in two test lungs as function of the different ventilator settings in experiment 2
Pplat, tPEEP and volume as a function of the different ventilator settings | |||||||
---|---|---|---|---|---|---|---|
1 Lung | |||||||
Vent setting | 1 | 2 | 3 | 4 | 5 | ||
Pplat | 28 | 28 | 26 | 28 | 36 | cmH2O | |
PEEPtot | 6 | 8 | 5 | 12 | 13 | cmH2O | |
Volume | 490 | 450 | 520 | 430 | 260 | ml | |
2 Lungs | |||||||
Traject A | |||||||
Vent setting | 1 | 2 | 3 | 4 | 5 | ||
Pplat | 28,56 | 28,56 | 25,84 | 35,36 | 46,24 | cmH2O | |
PEEPtot | 5,44 | 8,16 | 4,08 | 9,52 | 16,32 | cmH2O | |
Volume | 520 | 360 | 520 | 260 | 300 | ml | |
compliance | 23 | 20 | 27 | 10 | 8 | ||
Traject B | |||||||
Vent setting | 1 | 2 | 3 | 4 | 5 | ||
Pplat | 28,56 | 27,2 | 25,84 | 36,72 | 46,24 | cmH2O | |
PEEPtot | 5,44 | 8,16 | 5,44 | 10,88 | 16,32 | cmH2O | |
Volume | 510 | 360 | 510 | 240 | 300 | ml | |
compliance | 22 | 21 | 27 | 12 | 9 |