Table 3.

Illustration that PEEP can be applied and individualized. Experiment set-up is illustrated in supplemental Fig. 3. The in-line PEEP in the expiratory tract of circuit A was the variable in this experiment, the Vent ePEEP represents the asked PEEP from the ventilator. PEEPex is the theoretical ePEEP from the ventilator and in-line PEEP. PEEPtot is the measured PEEP in circuit A. PEEPint is the calculated intrinsic PEEP. All values are cmH2O unless specified otherwise.

Setting 1
Vent ePEEP 5 5 8 0
in-line PEEP (circuit A) No Yes No Yes
Circuit A
PEEPex 5 12.5 8 7.5
PEEPtot 6.8 13.6 8.2 9.5
PEEPint 1.8 1.1 0.2 2
Volume (mL) 490 250 480 370
Circuit B
PEEPex 5 5 8 0
PEEPtot 5.4 6.8 8.1 2.7
PEEPint 0.4 1.8 0.1 2.7
Volume (mL) 490 490 460 510
Setting 5
Vent ePEEP 15 15 8 0
in-line PEEP (circuit A) No Yes No Yes
Circuit A
PEEPex 15 22.5 8 22.5
PEEPtot 12 18 10 10
PEEPint − 3 − 4.5 2 − 12.5
Volume (mL) 160 120 300 200
Circuit B
PEEPex 15 15 8 0
PEEPtot 12.2 12.2 8.1 2.7
PEEPint − 2.8 − 2.8 0.1 2.7
Volume (mL) 190 190 280 280