Table 1.

Characteristics of available continuous glucose monitoring (CGM) systems in Korea

Model GuardianTM Connect CGM system Dexcom G5® Mobile CGM system Dexcom G6® CGM system FreeStyle® Libre iPro2 Professional CGM
Manufacturer Medtronic Dexcom Inc. Dexcom Inc. Abbott Diabetes Care Medtronic
Type Real time Real time Real time Intermittently scanned Retrospective
Age All ages ≥2 Years ≥2 Years ≥4 Years All ages
Sensor duration 6 Days (Enlite sensor) 7 Days 10 Days 14 Days 6 Days
7 Days (Guardian 3 sensor)
Calibration Every 12 hours Every 12 hours No (factory-calibrated) No (factory-calibrated) Every 6 hours
MARD Enlite sensor: 13.6% 9%–10% Overall 9% 11.40% 11.0%–12.2%
Guardian 3 sensor: 8.7%–10.5%
Warranty of transmitter 1 Year 3 Months 3 Months - -
Acetaminophen interference Yes Yes No No Yes
Reimbursement Yes Yes No* No* No
Software CareLinkTM Dexcom CLARITY® Dexcom CLARITY® LibreView® CareLinkTM iPro

MARD, mean absolute relative difference.

*

As of September 2020, FreeStyle Libre sensor and Dexcom G6 sensor are not reimbursed.