. 2020 Aug 3;103(4):1549–1552. doi: 10.4269/ajtmh.20-0212

Table 2.

Performance evaluation of microscopy and the Loopamp Malaria Pv Detection Kit for Plasmodium vivax identification, using PCR as the reference method

Statistic Microscopy Pv loop-mediated isothermal amplification
Plasmodium vivax infections True negative 379 351
True positive 139 151
False positive 1 29
False negative 40 28
Sensitivity (95% CI) 77.7 (71.6–83.8) 84.4 (79.0–89.7)
Specificity (95% CI) 99.7 (99.2–100) 92.4 (89.7–95.0)
Positive predictive value (95% CI) 99.3 (99.7–100) 83.9 (78.5–89.3)
negative predictive value (95% CI) 90.5 (87.6–93.3) 92.6 (90.0–95.2)
Cohen’s kappa 0.8212 SE 0.0265 (0.7693–0.8731) 0.7661 SE 0.0292 (0.7088–0.8234)

LAMP = loop-mediated isothermal amplification; SE = standard error.