Skip to main content
. 2020 Oct 9;101:382–390. doi: 10.1016/j.ijid.2020.10.008

Table 2.

Sensitivity by lateral flow test and immunoglobulin class detected.

Test IgM
IgG
IgM or IgG
n Positive Sensitivity (95% CI) n Positive Sensitivity (95% CI) n Positive Sensitivity (95% CI)
One Step COVID-2019 Test (Guangzhou Wondfo Biotech) 173 124 71.7 (64.3–78.2)
TR DPP® COVID-19 IGM/IGG Bio-Manguinhos (Fundação Oswaldo Cruz) 173 91 52.6 (45.2–59.9) 173 109 63 (55.6–69.8) 173 119 68.8 (61.5–75.2)
COVID-19 ECO IGM/IGG (Eco Diagnostica) 173 125 72.3 (65.2–78.4) 173 114 65.9 (58.6–72.6) 173 130 75.1 (68.2–80.9)
COVID-19 IgG/IgM (Qingdao Hightop Biotech) 173 39 22.5 (16.9–29.3) 173 89 51.4 (44.0–58.7) 173 91 52.6 (45.2–59.9)
Imuno-Rápido COVID-19 IgG/IgM (Wama Produtos Para Laboratorio) 173 70 40.5 (33.5–47.9) 173 108 62.4 (54.9–69.3) 173 119 68.8 (61.6–75.2)
COVID-19 IgG IgM (Gold Analisa Diagnóstica) 173 23 13.3 (9.1–19.2) 173 98 56.6 (49.2–63.8) 173 100 57.8 (50.4–64.9)