Table 1.

bCPAP design variations

Interface resistance Dead space Expiratory tubing resistance Comment
Sahni and Wung5 Low Low Low Used in several trials and manuscripts; Fisher & Paykel bCPAP has a similar design
Audu et al 8 High Low High Description of modified oxygen cannula (figures 1–3) later used in clinical trial
Brown et al 31 Low High Low* Description of the first version of the Pumani CPAP (figure 1)
Daga et al 9 High Low High Clinical trial using modified oxygen cannula with seal (figures 1–5)
Duke†10 High Low High Review including modified oxygen cannula (figures 2 and 3). Other designs: original bCPAP (figure 1) and high-resistance interface (RAM type, figure 5).
Kawaza et al 32 Low High Low* Clinical trial including Pumani first version, reference to Brown et al
Chisti et al11 High Low High Randomised trial including modified oxygen cannula, reference to Duke
McAdams et al 23 High Low Unknown Case series and description of bCPAP with high-resistance interface (RAM type) (figure 2)
Ezenwa et al 12 High Low High Retrospective observational study using modified oxygen cannula (figures 1 and 2)
WHO†13 High Low High Manual including modified oxygen cannula (figures 15 and 16). Other designs: original CPAP (figure 14).
Bennett et al 33 High High Low* Description of a bCPAP system with a high-resistance interface (RAM type) and connector tube similar to Brown et al (figure 1)
Falk et al 30 Low Low High Description of the second version of the Pumani CPAP (figure 2)
Bjorklund et al 39 High Low High Clinical trial using modified oxygen cannula with seal (figure 1)
Thaddanee et al 14 High Unknown High Observational trial and description of two modified oxygen cannula systems (figures 1 and 2).
Amadi et al 15 High Low High Observational trial and description of modified oxygen cannula systems (figure 1)

List of manuscripts with descriptions or use of bCPAP systems with design alterations from the original (top row). Dimensions are not always reported in manuscripts and some include several design variations. The figures or photographs refered to in the comment column can be found in the cited manuscripts.

*Data in the manuscript indicate that expiratory tubing resistance was not high.

†Properties refer to only one system in the manuscript. Details are provided in comment column.

bCPAP, bubble continuous positive airway pressure.