Table 3.

Sensitivity of the multiplex real-time PCR assay.

Pathogens Concentration Total Positive Detection rate 95% detection rate
PEDV 1000 copies/μL 23 23 100.0% > 95%
100 copies/μL 23 23 100.0% > 95%
10 copies/μL 23 20 87.0% < 95%
PDCoV 1000 copies/μL 23 23 100.0% > 95%
100 copies/μL 23 23 100.0% > 95%
10 copies/μL 23 14 60.9% < 95%
PToV 1000 copies/μL 23 23 100.0% > 95%
100 copies/μL 23 23 100.0% > 95%
10 copies/μL 23 17 73.9% < 95%
SADS-CoV 1000 copies/μL 23 23 100.0% > 95%
100 copies/μL 23 23 100.0% > 95%
10 copies/μL 23 12 52.2% < 95%

The cutoff line of positivity is automatically decided by Roche LightCycler® 96 Instrument.