Skip to main content
. 2020 Sep 29;7:583404. doi: 10.3389/fvets.2020.583404

Table 2.

Method validation data for canine plasma LC-MS/MS assay.

CBD 6-OH-CBD 7-OH-CBD THC 11-OH-THC THC-COOH CBC
Internal standard CBD-D3 7-OH-CBD-D9 7-OH-CBD-D9 THC-D3 11-OH-THC-D3 THC-COOH-D3 CBC-D9
Standard curve (ng/mL) 1.97–250 1.97–250 3.91–250 1.97–250 1.97–250 1.97–250 1.97–250
QC range (ng/mL) 5–175 5–175 10–175 5–175 5–175 5–175 5–175
LLOQ (ng/mL) 1.97 1.97 3.91 1.97 1.97 1.97 1.97
LOD (ng/mL) 0.98 0.98 0.98 0.98 0.98 0.98 0.98
Average cannabinoidrecovery (%) 91.15 79.43 87.11 95.24 92.51 93.60 86.62
Intraday accuracy (%) LQC = 113.4
MQC = 103.5
HQC = 103.8
LQC = 107.7
MQC = 106.9
HQC = 103.5
LQC = 107.5
MQC = 103.2
HQC = 99.9
LQC = 113.5
MQC = 103.7
HQC = 105.6
LQC = 109.8
MQC = 106.8
HQC = 105.4
LQC = 112.9
MQC = 105.2
HQC = 104.2
LQC = 113.5
MQC = 105.7
HQC = 106.6
Intraday precision (%) LQC = 0.32
MQC = 3.8
HQC = 2.1
LQC = 4.7
MQC = 3.6
HQC = 1.7
LQC = 3.2
MQC = 1.7
HQC = 1.4
LQC = 2.1
MQC = 2.4
HQC = 1.0
LQC = 5.5
MQC = 3.3
HQC = 2.6
LQC = 1.1
MQC = 3.8
HQC = 1.5
LQC = 1.2
MQC = 4.4
HQC = 2.5
Interday accuracy (%) LQC = 113.8
MQC = 103.9
HQC = 104.0
LQC = 108.1
MQC = 104.2
HQC = 100.4
LQC = 104.5
MQC = 102.6
HQC = 99.0
LQC = 111.4
MQC = 102.3
HQC = 103.1
LQC = 110.0
MQC = 104.0
HQC = 104.3
LQC = 113.7
MQC = 103.6
HQC = 103.1
LQC = 113.1
MQC = 103.9
HQC = 104.1
Interday precision (%) LQC = 0.56
MQC = 4.5
HQC = 2.7
LQC = 2.9
MQC = 2.7
HQC = 2.6
LQC = 4.6
MQC = 2.2
HQC = 2.4
LQC = 2.7
MQC = 2.9
HQC = 1.8
LQC = 4.5
MQC = 4.2
HQC = 2.7
LQC = 1.9
MQC = 3.1
HQC = 2.2
LQC = 1.1
MQC = 2.9
HQC = 1.7

LQC, low-quality control (5 ng/mL for CBD, CBC, 11-OH-THC, THC-COOH, and 6-OH-CBD; 10 ng/mL for THC and 7-OH-CBD); MQC, medium-quality control (100 ng/mL); HQC, high-quality control (175 ng/mL).