. 2020 Aug 24;12(9):928. doi: 10.3390/v12090928

Table 2.

Evaluation of assay performance in microtiter plates.

Plate
Format
Sample Ct Value HBV Quantification (Copies/mL) %CV 1 Z’ 2 Window 3
DMSO LMV DMSO LMV DMSO LMV
96 Intra 14.20 ± 0.08 23.41 ± 0.30 9.47 × 1010 ± 3.95 × 109 4.10 × 108 ± 9.76 × 107 2.56 0.91 0.88 230
Extra 20.54 ± 0.17 28.06 ± 0.21 1.31 × 107 ± 1.77 × 106 4.71 × 104 ± 6.80 × 103 0.87 0.76 0.85 278
384 Intra 15.07 ± 0.35 21.33 ± 0.63 2.31 × 1010 ± 4.92 × 109 7.58 × 108 ± 1.88 × 108 2.34 2.96 0.53 30
Extra 31.60 ± 0.80 44.80 ± 0.41 3.53 × 106 ± 1.33 × 105 2.80 × 103 ± 6.50 × 102 2.56 0.92 0.72 1260

1 %CV (coefficient of variation). 2 Z’ determined as described by Zhang et al., 1999 [9]. 0.5 ≤ Z’ < 1 indicates “high quality” of the assay. 3 Window (signal-to-background ratio).