Skip to main content
. 2020 Oct 12;19:77. doi: 10.1186/s12938-020-00821-z

Table 1.

Control peak temperature and lesion results as a function of RF power for Study 1 and Study 2

Power (W) Study Esophageal lesion depth (mm) Maximum esophageal fraction of damage (%) Peak temperature (°C)
10 Control 1 0.00 1.14 50.07
Control 2 0.52 3.22 50.11
20 Control 1 0.77 4.56 65.25
Control 2 1.88 34.71 65.29
30 Control 1 1.72 24.68 82.56
Control 2 2.50 100.00 82.96
40 Control 1 2.44 77.44 99.63
Control 2 2.50 100.00 100.85
50 Control 1 2.50 100.00 111.31
Control 2 2.50 100.00 112.76

Numerical values are from Fig. 1