Table 4.

Intraobserver reliability analysis of patellar tendon stiffness in seventy-six athletes with patellar tendinopathy (N = 76 tendons)

Analysis 1 (SB) Analysis 2 (SB) Intraobserver reliability
Location Stiffness (kPa) Stiffness (kPa) CV (%)a CR (kPa)b ICC (95%CI)c
All ROIs 74.9 [56.4–105.4] 69.9 [54.7–100.3] 10.3 (7.9–12.2) 11.9 [5.1–24.9] 0.95 (0.92–0.97)
ROI 1 78.3 [51.6–117.3] 78.4 [51.7–111.6] 15.8 (9.2–20.3) 14.8 [5.3–26.8] 0.92 (0.88–0.95)
ROI 2 85.4 [55.9–127.6] 72.4 [54.4–111.8] 15.1 (12.3–17.4) 16.2 [7.0–44.4] 0.89 (0.83–0.93)
ROI 3 69.7 [52.7–102.3] 63.8 [49.2–84.5] 19.2 (13.3–23.6) 20.2 [6.8–41.6] 0.85 (0.77–0.90)
ROI 4 59.9 [41.7–76.3] 48.9 [34.1–61.9] 18.9 (11.2–24.3) 13.3 [5.5–24.3] 0.92 (0.85–0.95)

Patellar tendon stiffness was assessed using shear-wave elastography (SWE), expressed as median (interquartile range) in kPa

aCV: coefficient-of-variation (%), 95% confidence interval

bCR: coefficient-of-repeatability (kPa), also referred to as the smallest real difference (SRD)

cICC: intraclass correlation coefficient (ICC), 95% confidence interval