Table 4.

Data related to the LOD, LOQ, slope, and interception for curcumin and quercetin HPLC analysis method.

Parameters Validation Results
Curcumin Linearity Calibration range (µg/mL): 0.5–12.5
y = 40,565x − 272.07 (R2 = 0.9998)
LOD 0.005 µg/mL
LOQ 0.017 µg/mL
Slope 40,565 ± 201
Intercept −272.07 ± 1246.54
Quercetin Linearity Calibration range (µg/mL): 0.5–12.5
y = 18,043x + 129.07 (R2 = 0.9997)
LOD 0.14 µg/mL
LOQ 0.48 µg/mL
Slope 18,043 ± 118
Intercept 129.07 ± 733.51