. 2020 Aug 27;10(9):202. doi: 10.3390/membranes10090202

Figure 4.

Figure 4

Chronoamperometric response of a PVC-membrane-coated glassy carbon (GC) electrode upon spiking the phosphate buffered saline (PBS) background electrolyte with propofol standard solutions. The concentrations in the PBS solution labeled 1 to 9 are 1.25, 2.5, 4.98, 9.9, 19.6, 38.4, 56.6, 80.5 and 111.1 μmol/L. The inset in the lower right is the calibration curve constructed from the chronoamperometric currents. The inset in the upper left corner shows linear sweep voltammograms recorded in PBS background solution and in PBS with propofol concentrations between 9.9 and 111.1 μmol/L. Membrane composition 25.5% PVC, 50.9% 2-nitrophenyl octyl ether (o-NPOE), 21.2% tetradodecylammonium tetrakis(pentafluorophenyl) borate (TDDATPFPhB), 2.4% sodium tetrakis[3,5bis(trifluoromethyl) phenyl] borate (NaTFPhB).