. 2020 Oct 15;11:5208. doi: 10.1038/s41467-020-18952-1

Table 2.

Comparison between normal dose exposures versus optimized dose exposures using typical high-resolution microscope systems and parameters.

Exposures/h with 1 exposure/hole (K2) Exposures/h with 4 exposures/hole (K2) Exposures/h with 1 exposure/hole (K3) Exposures/h with 4 exposures/hole (K3)
Normal dose (~66 e-/Å2) 108 121 195 237
Optimized dose (~17 e-/Å2) 178 190 242 273
% increase in collection efficiency 65% 57% 25% 15%

Each entry reports the number of exposures collected using the normal versus denoised optimized doses using either a K2 or K3 camera on a Titan Krios and collecting only one exposure per hole versus four exposures per hole. The optimized dose yields more than a 50% increase in throughput on the K2 camera and still yields a modest improvement in throughput on the newer K3 camera.