Skip to main content
. Author manuscript; available in PMC: 2020 Oct 16.
Published in final edited form as: Med Phys. 2020 May 23;47(8):3263–3276. doi: 10.1002/mp.14201

TABLE II.

Quantitative evaluation of the segmentation accuracy.

Patient Number of BMs COMS/mm HD/mm MSSD/mm SDSSD/mm Pre-FNR Pre-FPoU Post-FNR Post-FPoU
1 19 0.89 1.86 0.60 0.56 0.21 0.32 0.21 0.17
2 12 1.47 2.67 0.90 0.96 0.08 0.70 0.08 0.25
3 23 1.53 3.20 1.39 0.77 0.04 0.68 0.04 0.34
4 15 1.84 3.99 0.99 1.11 0.00 0.63 0.00 0.29
5 40 1.62 2.97 0.96 0.76 0.08 0.42 0.08 0.22
6 16 1.49 3.42 1.57 0.81 0.13 0.47 0.13 0.30
7 30 1.01 2.13 0.67 0.59 0.20 0.30 0.20 0.09
8 54 1.98 3.42 1.34 0.87 0.22 0.17 0.22 0.08
9 50 1.91 3.22 1.09 0.82 0.30 0.34 0.30 0.12
10 81 1.72 2.96 1.13 0.79 0.26 0.23 0.26 0.07
Mean(SD) 34 ± 22 1.55 ± 0.36 2.98 ± 0.63 1.06 ± 0.31 0.80 ± 0.16 0.15 ± 0.10 0.43 ± 0.19 0.15 ± 0.10 0.19 ± 0.10