Table 3.

Left ventricular parameters in fully automated and expert corrected analysis

Fully automated Expert corrected P value
Median IQR Median IQR
Absolute values
 LV EDV (ml) 160.9 124.6–211.8 162.4 126.7–213.1 0.0004
 LV ESV (ml) 73.1 48.5–121.3 66.1 46.3–118.9 0.0008
 LV SV (ml) 78.8 59.4–95.0 83.7 68.7–98.6  < 0.0001
 LV EF (%) 52.2 41.0–62.2 56.0 42.6–62.7  < 0.0001
 LV mass (g) 166.5 134.4–194.8 165.5 133.0–195.8 0.7677
Indexed to body surface area
 LV EDV (ml/m2) 78.3 63.4–98.3 79.2 65.6–100.2 0.0005
 LV ESV (ml/m2) 35.1 25.7–59.5 32.5 25.1–59.0 0.0005
 LV SV (ml/m2) 39.8 28.4–45.5 41.4 34.0–49.9  < 0.0001
 LV mass (g/m2) 81.1 64.3–91.9 80.7 66.4–89.8 0.7998

P value is for comparison of fully automated and expert corrected results using Wilcoxon matched pairs test