Table 1.

Summary of the performance and analytical considerations from different hybrid approaches.

Instrumentation Biological Sample Method Validation Metabolites Coverage and RT (min) Data Processing Strategies Ref
UHPLC-Q-TOF-MS and UHPLC-QTRAP-MS Human
urine
69% and 94% of metabolites displayed a relative standard deviation (RSD) of <10% and <20%, respectively. After normalization and internal standard calibration, 94% and 97% of metabolites had an RSD of < 20%, respectively. 419 compounds in electrospray ionization quadrupole (ESI) positive mode and 449 compounds in ESI negative mode were detected.
RT = 26 min.
Analyst 1.6 software (AB SCIEX, USA) was used to conduct peak integration. The non-targeted analysis data were imported to the SIEVE software (ThermoFisher, USA) package to extract the metabolite features. [40]
LC-QTRAP-MS Human
urine
The slopes of linear regressions were approximately 1.00 (1.0291). RSDs were less than 9.0% with a correlation coefficient (R2) of 0.9998. The number of accurately quantified metabolites from 103 thiols, increased from 64 (62%) to 99 (96%).
RT = 55 min.
Manual data acquisition and processing through Bruker Daltonics Control 3.4 and Bruker Daltonics Data analysis 4.0 software (Bruker Daltonics, Bremen, Germany). [41]
GC-MS Human
urine
75.7%, 89.5%, and 95.9% of the peaks had RSDs of <10%, 20%, and 30%, respectively. The intraday RSDs for 80.2%, 91.6%, and 95.1% of the peaks were <10%, 20%, and 30%, respectively. The interday RSD for 64.3%, 86.9%, and 91.6% of the peaks were <10%, 20%, and 30%, respectively. A total of 76 differential metabolites were defined, 58 of which were verified.
RT = 30 min.
Changes in the levels of the differential metabolites were visualized with MultiExperiment Viewer (http://www.tm4.org). The altered pathways were determined with MetaboAnalyst 2.0 (http://www.metaboanalyst.ca). [52]
UHPLC-Q-Orbitrap-MS and UHPLC-QTRAP-MS Human
urine
The RSDs of all the standards were below 10%. 94% and 80% of the peaks had RSDs of <30% and <20% in the QC samples, respectively. The 95.5% and 90% peaks had RSDs of <30% and <20% in the urine samples, respectively. 780 metabolites were defined. A total of 48 metabolites were chosen, 26 of which were identified.
RT = 28 min.
Ion-pairs selection was from two sources: (1) untargeted metabolic analysis; (2) key metabolites in the metabolic pathways chosen from free databases. MultiQuant software (Applied Biosystem/MDS Sciex, Carlsbad, CA) was used to extract the peaks. [53]
UHPLC-Q-TOF-MS and UHPLC-QQQ-MS Human
serum
34% of the detected metabolites had an RSD < 5% and 76% had an RSD < 10%. 518 ion-pairs were defined for subsequent MRM detection.
RT = 29.9 min.
Manual peak detection and alignment through XCMS software. [37]
LC-QTOF-MS Human
serum
The minimum absolute height required was set at 3000 counts, which was also used for 100% of analyses in the recursive step to minimize the experimental variability. 139 and 158 molecular entities in the negative and positive ionization mode, respectively, were obtained.
RT = 20 min.
The MassHunter Workstation software package was used to process all the data obtained by LC/Q-TOF in the MS/MS mode. [38]
UHPLC-Q-TOF-MS and UHPLC-QQQ-MS Human
serum
91% of contents had an RSD of <30% from the QC samples. 1446 metabolite MRM transitions in the MRM-Ion Pair Finder.
RT = 23 min.
Efficient data processing strategy, with the processing time markedly shortened by the homemade MRM-Ion Pair Finder software. [31]
LC−QQQ-MS Human
serum
The linearity r at 0.82 ± 0.26 of amino acids. >40% of the GOT-MS MRMs had CVs < 5%; the average CV of the detected metabolites was at 7.8 ± 7.0%. 595 precursor ions and 1890 multiple reaction monitoring transitions (MRM).
RT = 9 min.
Manual inspection based on symmetry, peak width, and MS peak area extraction by Agilent MassHunter Qualitative Analysis and Quantitative Analysis software (Agilent Technologies, Inc., Santa Clara, CA). [24]
UHPLC-Q-TOF-MS and UHPLC/QTRAP-MS Human
serum
The change folds of a peak area between these two QC samples ranged from 0.002 to 14. 767 (94.3%) and 759 (93.3%) of the peaks had a CV of <30% in the b-QC and p-QC replicates, respectively. 813 ions were steadily detected in the QC samples.
RT = 30 min.
The integrations of the peak areas were processed by the software provided by the instrument vendor. The efficient data processing strategy used the MRM-Ion Pair Finder software. [28]
UHPLC-QTRAP-MS and LC-MS Human serum For the validation set, the AUC was from 0.676 to 0.875. Sensitivity was from 0.504 to 0.921. Specificity was from 0.528 to 0.784. 239 metabolites were identified.
RT = 20 min.
Manual peak detection and alignment through the XCMS software. Home-developed database and online databases (HMDB and Metlin) or confirming with authentic standards were used. [54]
2 D-UHPLC-QTRAP-MS Human
plasma
The linear regression r varied between 0.9902 and 0.9993. The average accuracies for the standard samples were between 0.11% and 13.90%, and the average intra-/inter-day precisions of the standard samples were between 0.66% and 16.46%. The intra-/inter-day precisions in the complex plasma matrix were between 2.05% and 19.49%. 78 metabolites were confidently confirmed, from which 73 metabolites can be accurately quantified. Untargeted profiling of 4651 features of high reliability and validity was achieved.
RT = 27 min.
Targeted qualitative and quantitative was performed using the TraceFinder 3.3 software (Thermo Fisher Scientific, Waltham, MA). All the calibration curves were linear and weighted 1/x. The untargeted analysis was performed by the Progenesis QI 2.0 software (Waters, Milford, MA, USA). [26]
UHPLC-Q-Orbitrap-MS and UHPLC-Qtrap-MS Human
plasma
43 yielded good linear functions (R2 > 0.99) had RSDs lower than 20% and an accuracy between 80% and 120%. 1658 characteristic ion-pairs from 1324 metabolites.
RT = 35 min.
48 metabolites established DITQM; manually generate “one feature for one peak” metabolomics data, further confirmed through the extracted ion chromatograms (XICs). [39]
UHPLC-Q-TOF-MS and UHPLC-QTRAP-MS Human plasma The intraday RSDs of 82.4% of the metabolites were <15%. 25.1% of the metabolites had RSD values of >15%. A total of 1373 unique metabolite ion-pairs were obtained in the positive ion mode. 162 significantly changed metabolites were defined.
RT = 30 min.
Peak detection and alignment were performed by the MarkerView software (AB SCIEX, Framingham, USA) MS-DIAL software and homemade C-package were used to handle MS2. UHPLC-MRM MS data were disposed of in Analyst 1.6 software (AB SCIEX, Framingham, USA). [55]
UHPLC-QQQ-MS Human plasma The optimized MRMs had intensities of >1000 and signal-to-noise-ratios (S/Ns) of >3. The intraday and interday median CVs were 4.86% and 8.79%, respectively. The median r was 0.96. 927 metabolites were measured. 310 were confirmed using pure chemical standards, while the rest were annotated by identification level using database entries.
RT = 15 min.
The entire LC-MS system was controlled by the Mass Hunter Workstation software (Agilent, Santa Clara, CA). The extracted MRM peaks were integrated using the Mass Hunter Quantitative Data Analysis software. [51]
UHPLC-Q-TOF-MS Mouse
serum
66.7% of metabolites displayed an RSD of <10%. 99.8% of metabolites had an RSD of <20%. After normalization by the total intensity and sample median, 99.8% and 99.4% of metabolites exhibited an RSD of <20%, respectively 2081 ions were obtained after data filtering. 569 peaks were selected to perform tsMIM-based pseudotargeted analysis.
RT = 26 min.
The raw data were extracted and aligned by the Progenesis QI software. The peak areas were collected by the TargetLynx software. [43]
UHPLC-Q-TOF-MS Mouse
plasma
97.4% and 95.4% of the selected ions in the positive and negative ion modes have RSD values of less than 20%, respectively. 1423 and 1141 ions were generated in the positive and negative ion modes, respectively.
RT = 27 min.
The raw data files were uploaded into the Progenesis QI software to perform chromatographic peak alignment, data normalization, and peak picking using the default settings. MassLynx XS (Waters Corp., Manchester, UK) was used for the targeted extraction. The selected features were tentatively annotated. [42]
UHPLCQ-TOF-MS Mouse
plasma
Almost 90% of the tsMIM- detected metabolites had an RSD of <20%. 42% of metabolites had RDSs of <5%. 387 ions were detected. 17 metabolites were selected as biomarkers.
RT = 26 min.
The peak areas of the metabolites were processed using the TargetLynx software. [27]
UHPLC-Q-Orbitrap-MS and UHPLC-QQQ MS Mouse brain and liver, cancer cells, and human plasma The linear of IS r > 0.99. The intraday RSDs were from 0.8% to 4.3%; the inter-day RSDs were from 2.4% to 16.8%. 51% and 94% of the detected lipids had RSDs of <5% and 20% in positive mode and 94% had RSDs of <10% in negative mode. A total of 3377 targeted lipid ion pairs with over 7000 lipid molecular structures were defined.
RT = 20 min.
The raw data were processed with Analyst software. Normalization of the lipids by the appropriate lipid ISs. [44]
UHPLC-Q-TOF-MS and UHPLC-QTRAP-MS Rice seed 89.5% of peaks had RSDs < 20%. The linear range was 2.5–4 orders of magnitude, and the r was in the range 0.996–0.999. The recoveries were 85.9–106.3% for positive mode and 73.3–98.2% for negative mode, respectively. A total of 749 and 617 ion pairs in the positive and negative modes were achieved, respectively. Among them, about 200 metabolites were identified or tentatively identified.
RT = 30 min.
All the ions were extracted by the Analyst software. Zero values were removed by the 80% rule. [28]
GC-MS Commercial
cigarettes
>81.2% of peaks had RSDs of <20%. 312 components and their related quantitation ions. A total of 90 compounds were elucidated.
RT = 72.5 min.
The integration of the chromatography peaks was performed using the Agilent MSD ChemStation (Agilent Technologies). Peak areas of all the components were divided by those of the internal standard and then scaled to zero mean and unit variance. [36]
GC-MS Tobacco leaf 47.3% of components had an r of >0.99; 36% of components had an RSD of <5%; 93% of components had an RSD of < 20%. 167 differential components (p < 0.05) were screened out.
RT = 72.5 min.
The acquired GC/MS raw data of the QC sample were imported into the AMDIS software (version 2.62, NIST, USA) for peak deconvolution and detection. The quantitative ion selection algorithm was written in Visual C++ (version 6.0, Microsoft). The component peak area was divided by that of the internal standard. [25]
SESI-QQQ-MS Bacterial
culture
47% of features showed an r of >0.90. 65.9% of features had a CV of <20% 75 features in the SESI-GOT-MS/MS panel were established.
RT = 1 min.
Manual inspection based on symmetry, peak width, and MS peak area extraction by Agilent MassHunter Qualitative Analysis and Quantitative Analysis software. [29]
UHPLC-QQQ-MS Bacterial
culture
54.9% of peaks had measurement CVs of <10%; 0.84% had CVs of >40%. 51.97% of peaks had CVs of <10%; 0% had CVs of >40%. A total of 464 metabolite peaks were detected.
RT = 20 min.
All the raw metabolomics data were inspected by the Quanbrowser module of Xcalibur version 4.0 (Thermo Fisher Scientific) and the Thermo TSQ LC-SIM Data Processor (V0.1.35 Engineering Sample). [45]