Table 2.

Accuracy and precision results for OTN determination by the proposed LC-MS/MS method.

Accuracy and Precision (n = 5, Three days) LLOQ (25 nM) MQC (500 nM) HQC (800 nM)
1st day Intraday Bias (mean, n = 5) 15.41 −6.30 −1.09
Average Percent Recovery 115.41% 93.70% 98.91%
S.D. 14.55 2.01 3.57
% R.S.D 12.61 2.14 3.64
2nd day Intraday Bias (mean, n = 5) 10.34 1.49 −0.10
Average Percent Recovery 110.34% 101.49% 99.90%
S.D. 15.37 1.90 1.82
% R.S.D 13.92 1.87 1.82
3rd day Intraday Bias (mean, n = 5) 13.80 0.98 −0.91
Average Percent Recovery 113.80% 100.98% 99.09%
S.D. 9.57 1.83 2.62
% R.S.D 8.41 1.81 2.64
Interday Bias (mean, n = 15) 13.40 −1.30 −1.01
Average Percent Recovery 113.40% 98.70% 98.99%
S.D. 13.41 4.02 2.89
% R.S.D 11.83 4.07 2.92