Table 3.

Comparison of interstitial glucose measured by FSL-CGM and reference venous blood glucose based on standard analytical approaches

Fasting (n = 17) 1 h post-OGTT (n = 17) 2 h post-OGTT (n = 17) All (n = 51)
Time difference between paired samples ± SD (minutes) 8.8 ± 5.4 5.3 ± 5.0 7.9 ± 5.2 7.4 ± 5.3
Correlation coefficient (95% CI) 0.57 (0.12–0.83) 0.68 (0.30–0.88) 0.66 (0.26–0.87) 0.81 (0.69–0.89)
MAD overall (mmol/L) −0.54 − 0.92 −0.89 − 0.78
MAD when ref. < 5.5 mmol/L (n = 21) −0.58
MARD overall (%) −12.2 −11.3 −13.2 −12.3
MARD when ref. > 5.5 mmol/L (n = 30) −11.9
Within ISO 2013 criteria (%) 58.8 41.2 47.1 49.0

SD standard deviation; CI confidence interval; MAD mean absolute difference; MARD mean absolute relative difference; ISO international organization of standardization