Skip to main content
. 2020 Aug 12;12(556):eabc7075. doi: 10.1126/scitranslmed.abc7075

Table 1. Shown is RT-qPCR and RT-LAMP testing of 768 clinical samples stratified into CT value bins (see Fig. 3A).

Fig. 3B and table S2 show specificity and sensitivity values calculated from these numbers.

RT-LAMP
CT Pos Neg Sum
RT-qPCR Pos 0–25 51 0 51
25–30 28 2 30
30–35 4 16 20
35–40 0 16 16
Neg Neg 2 649 651