Skip to main content
. 2020 Oct;15(5):783–791. doi: 10.26603/ijspt20200783

Table 3.

Intra-rater (test-retest), intra-day reliability for the handheld (HH) oximeter and the Doppler ultrasound

ICC2.1 [95% CI] SEM mmHg SDC95 mmHg
HH oximeter 0.72 [0.52 to 0.83] 10 27
Ultrasound 0.79 [0.63 to 0.88] 8 22

ICC: Intraclass Correlation Coefficient

SEM: Standard Error of Measurement

mmHg: Thigh cuff pressure measured in millimeters of mercury

CI: Confidence Interval

SDC95: Smallest Detectable Change with a 95% CI