Validation of volume-based qPCR method for transgene detection in human blood.

Nominal concentration (copies/μL blood) 1st day 2nd day 3rd day
Mean observed concentration (copies/μL blood) Relative error (%) Coefficient of variation (%) Mean observed concentration (copies/μL blood) Relative error (%) Coefficient of variation (%) Mean observed concentration (copies/μL blood) Relative error (%) Coefficient of variation (%)
200,000 218,066.2 9.0 5.4 196,127.5 − 1.9 6.9 190,866.7 − 4.6 9.0
2,000 1895.1 − 5.2 10.3 1803.6 − 9.8 7.4 1852.4 − 7.4 8.1
7.5 8.1 7.4 7.0 8.2 9.4 4.4 7.3 − 2.7 5.7
Slope − 3.35 − 3.39 − 3.33
intercept 8.35 8.25 8.37
PCR efficiency (%) 98.8 97.2 99.5
R2 1.000 0.999 1.000