Skip to main content
. Author manuscript; available in PMC: 2020 Oct 22.
Published in final edited form as: Metrologia. 2017;54(1A):https://doi.org/10.1088/0026-1394/54/1a/06011.

Table 5:

BIPM.RI(II)-K4. comparison results and link to the BIPM.RI(II)-K1 comparisons

Radionuclide Measurement method ACRONYM* Solution
volume
(calculated)
/cm3
AE/kBq u(AE)/kBq Linked Ae/kBq u(Ae)/kBq
18F IC calibrated 4P-IC-GR-00-00-00 in August 2015 by 4π(LS)β-γ anticoinc. (LTAC) 4P-LS-BP-NA-GR-AC 3.57 and 3.62 10.228 0.041 15 291 64
64Cu IC calibrated 4P-IC-GR-00-00-00 in March 2016 by 4π(LS)β-γ anticoinc. (lTAC) 4P-LS-BP-NA-GR-AC 3.57 and 3.58 54.34 0.32 80 540 500
*

See appendix 1