Skip to main content
. 2020 Oct 20;6(10):e05265. doi: 10.1016/j.heliyon.2020.e05265

Table 1.

Stable isotopic compositions including clumped isotope data of IAEA NBS-19 for reproducibility test. These samples were analysed during 2013–2014. The working gas (WG) used was AS-2 with δ13 C = -32.54 ‰ and δ18 O = -4.67 ‰ (in VPDB scale).

Sl. No. δ13 C (‰) (VPDB) δ18 O (‰) (VPDB) δ47 (WG) (‰) Std. Err. Δ47 (‰)
(ARF)
Std. Err
1 2.02 -2.21 35.22 0.01 0.382 0.010
2 2.02 -2.11 35.54 0.02 0.394 0.012
3 2.02 -2.19 35.28 0.01 0.416 0.010
4 2.01 -2.28 35.15 0.01 0.408 0.011
5 2.00 -2.27 35.24 0.02 0.388 0.016
6 2.00 -2.16 35.27 0.02 0.370 0.013
7 2.02 -2.27 35.21 0.01 0.398 0.009
8 2.02 -2.20 36.48 0.01 0.363 0.008
9 2.01 -2.20 36.56 0.02 0.392 0.006
10 2.01 -2.15 36.46 0.01 0.399 0.012
11 2.01 -2.20 36.57 0.01 0.393 0.010
12 2.02 -2.21 36.32 0.01 0.387 0.009
13 2.01 -2.18 36.43 0.01 0.368 0.014
14 2.01 -2.16 35.81 0.01 0.379 0.010
15 2.00 -2.18 35.76 0.01 0.387 0.006
Average 2.01 -2.20 35.82 0.388
Std. Dev. 0.01 0.05 0.58 0.014