Skip to main content
. 2020 Sep 27;20(19):5532. doi: 10.3390/s20195532

Table 2.

Analytical parameters of the calibration curves obtained from individual and multianalyte immunoassays on gold SPE 8x compared with the parameters previously evaluated by ELISA [11], fluorescent microarray [31] and individual SPE in 1× format [32].

ELISA [11] *1 Microarray [31] *1 SPE 1x Format [32] *2 SPE 8× Format *1 Multianalyte SPE 8× Format *2 ELISA [11] Microarray [31] Individual SPE 8× Format Multianalyte SPE 8× Format
Irgarol 1051 Sulfapyridine
Signalmax 100.4 ± 0.4 100.6 ± 2.1 95.42 ± 4.40 92.2 ± 6.0 97.3 ± 6.5 100.4 ± 0.6 97.1 ± 5.1 90.9 ± 10.0 91.3 ± 4.3
Signalmin 2.9 ± 0.4 3.1 ± 1.9 −2.45 ± 4.78 −0.6 ± 5.2 23.3 ± 8.3 2.7 ± 1.2 −2.5 ± 6.4 −0.22 ± 8.4 9.9 ± 7.0
Slope −0.98 ± 0.22 −1.45 ± 0.19 −0.81 ± 0.16 −1.07 ± 0.26 −5.01 ± 4.9 −0.784 ± 0.09 −1.14 ± 0.36 −0.93 ± 0.37 −3.50 ± 1.51
IC50, nM 0.58 ± 0.19 2.29 ± 1.11 4.12 ± 1.31 4.77 ± 1.27 19.89 ± 1.16 6.58 ± 1.73 15.23 ± 1.34 5.65 ± 1.57 13.85 ± 1.20
IC50 µg/L 0.145 ± 0.05 0.579 ± 0.28 1.04 ± 0.33 1.20 ± 0.23 5.04 ± 0.29 1.43 ± 0.23 3.79 ± 0.33 1.40 ± 0.39 3.45 ± 0.29
LOD
µg/L
0.012 ± 0.007 0.135 0.038 ± 0.022 0.037 3.24 0.08 ± 0.02 0.397 0.009 1.08
LOQ
µg/L
- - 0.24 ± 0.23 0.207 3.97 - - 0.16 2.05
R2 0.999 0.997 0.98 0.85 0.77 0.999 0.983 0.77 0.89
Chloramphenicol Estradiol
Signalmax 102.6 ± 8.8 98.8 ± 2.8 - 98.2 ± 3.8 97.2 ± 6.5 99.8 ± 0.5 97.7 ± 1.7 89.9 ± 6.7 96.1 ± 4.3
Signalmin 8.8 ± 1.2 28.7 ± 3.9 - 1.5 ± 3.4 20.0 ± 10.5 0.6 ± 0.7 −0.05 ± 2.39 0.45 ± 8.2 20.4 ± 6.5
Slope −0.60 ± 0.05 −1.18 ± 0.32 - −1.30 ± 0.19 −2.86 ± 1.99 −1.01 ± 0.07 −1.23 ± 0.15 −1.32 ± 0.62 −9.56 ± 4.3
IC50, nM 0.59 ± 0.05 10.25 ± 1.28 - 3.47 ± 1.15 12.91 ± 1.36 4.01 ± 0.74 9.67 ± 1.11 35.4 ± 1.36 21.83 ± 1.05
IC50 µg/L 0.192 ± 0.09 3.00 ± 0.41 - 1.12 ± 0.37 4.17 ± 0.44 1.09 ± 0.20 2.63 ± 0.30 9.72 ± 0.37 5.94 ± 0.28
LOD
µg/L
0.004 ± 0.01 0.453 - 0.18 1.89 0.11 ± 0.03 0.362 0.41 4.60
LOQ
µg/L
- - - 0.36 2.70 - - 1.93 5.18
R2 0.997 0.988 - 0.94 0.72 0.999 0.998 0.90 0.82

*1 PBST; *2 sea water.