Skip to main content
. 2020 Aug 27;34(11):799–806. doi: 10.1007/s12149-020-01512-4

Table 3.

Survey results and DRLs: hybrid CT of SPECT/CT

Target site, purpose of use CTDIvol, mGy DLP, mGy cm
50th percentile 75th percentile 2020 DRLs 50th percentile 75th percentile 2020 DRLs
Whole body, AC + Dx 2.9 5.03 5.0 180 384.1 380
Brain, AC + Dx 10.6 22.60 23.0 262 419.2 410
Head and neck, AC + Dx 3.3 5.73 5.8 133 205.1 210
Chest, AC + Dx 2.6 4.06 4.1 90 164.8 170
Heart, AC + Dx 3.2 4.50 4.5 89 180.0 180
Abdomen, AC + Dx 3.0 5.02 5.0 89 206.8 210
Extremities, AC + Dx 3.2 4.69 4.6 159 237.2 230
Brain, AC only 7.3 12.86 13.0 140 350.9 330
Heart, AC only 1.6 4.04 4.1 45 84.1 85