Skip to main content
. Author manuscript; available in PMC: 2021 Apr 26.
Published in final edited form as: JACC Cardiovasc Interv. 2020 Sep 30;13(20):2361–2370. doi: 10.1016/j.jcin.2020.06.047

TABLE 2.

Case Planning

Patient 1 Patient 2 Patient 3 Patient 4 Patient 5
Reason for surgical ineligibility Frailty Frailty Pulmonary fibrosis Third redo sternotomy 86 years of age
Septal-lateral diameter (mm) 33.4 36.9 29.3 27.9 31.9
Intercommissural diameter (mm) 41.5 47.7 42.3 43.2 45.4
Perimeter (mm) 126.4 141.3 122.9 116.7 128.9
Area (cm2) 11.5 13 10.5 9.7 12.1
Implanted valve size 35M LP 37L SP 33M LP 33M LP 37M LP
Septal-lateral percentage oversizing 3.3 −1.1 10.8 12.1 14.4
Intercommissural percentage oversizing 16.9 10.1 10.0 8.1 9.0
Perimeter percentage oversizing 13.9 10.4 17.1 13.3 16.4
Predicted neo-LVOT (mm2) 571 294 306 296 341
End-systolic aortomitral angle (deg) 32.2 NA NA 52.1 54.2
End-diastolic aortomitral angle (deg) 33.7 NA NA 53 49.3

Values were derived using contrast-enhanced gated cardiac computed tomography for the 5 patients undergoing electrosurgical laceration and stabilization of MitraClip.