TABLE 3.
Procedural and In-Hospital Outcomes (N = 5)
Successful electrosurgical laceration | 5 (100) |
Number of clips lacerated (per patient) | 2 |
Successful deployment and correct positioning of Tendyne valve | 5 (100) |
Need for valve retrieval | 0 |
Procedure time (min) | 205 (198.5–237.5) |
Fluoroscopy duration (min) | 59.5 (54.6–66.4) |
Planned intra-aortic balloon pump | 5 (100) |
Hospital length of stay (days) | 13 (11–37) |
Procedural complications | |
Death | 0 |
Cardiac arrest | 0 |
Tamponade | 0 |
Cardiac perforation | 0 |
LVOT obstruction | 0 |
Required conversion to open surgery | 0 |
Significant hemodynamic compromise | 1 (20) |
Laceration of nontarget structure | 0 |
Valve embolization | 0 |
Stroke | 0 |
Major vascular complication | 0 |
Life-threatening bleed | 2 (40) |
Hemothorax | 2 (40) |
Acute renal failure | 1 (20) |
Perivalvular leak greater than mild | 2 (40) |
Discharge echocardiographic outcomes | |
Left ventricular ejection fraction (%) | 55 (50.0–57.5) |
Mean mitral gradient (mm Hg) | 5.0 (4.3–5.5) |
Central MR grade 0 | 5 (100) |
Perivalvular leak greater than mild | 2 (40) |
Survival | |
Survived immediate procedure | 5 (100) |
Survived to discharge | 5 (100) |