Table 1. Specifications of the AO-FF-SS-OCT system.
Imaging source bandwidth | 825 nm to 875 nm |
Wavefront beacon bandwidth | 740 nm to 770 nm |
Lateral resolution (in the retina) | 2.6 µm |
Axial resolution (in the retina) | 7.3 µm |
Volume rate | 200 Hz to 1000 Hz |
Measured system sensitivity | 63 dB @ 200 Hz |
A-scan rate (effective) | 8.2 MHz to 18.4 MHz |
Sensitivity roll-off (6 dB) | 2.1 mm |