. 2020 Jun 11;23(4):575–583. doi: 10.1007/s40477-020-00481-8

Table 4.

Sensitivity, specificity, positive predictive value and negative predictive value of pre-specified elastography cutoffs for lesions located in the optimal field of view -5 to 40 mm (n=96)

Elastography cutoff valuea Sensitivity (95% CI) Specificity (95% CI) Predictive value
Positive Negative
36 kPa 50/54 30/42 50/62 30/34
92.6% 71.4% 80.6% 88.2%
(82.1–97.9) (55.4–84.3) (68.6–89.6) (72.5–96.7)
50 kPa 49/54 33/42 49/58 33/38
90.7% 78.6% 84.5% 86.8%
(79.7–96.9) (63.2–89.7) (72.6–92.6) (71.9–95.6)
80 kPa 40/54 40/42 40/42 40/54
74.1% 95.2% 95.2% 74.1%
(60.3–85) (83.8–99.4) (83.8–99.4) (60.3–85)

a The test was considered positive when the cutoff value was exceeded