Septal Reduction Using Transvenous Intramyocardial Cerclage Radiofrequency Ablation

Preclinical Feasibility

Eun-Seok Shin a, Min-Ku Chon b, Eun Jung Jun a, Yong-Hyun Park b, Sang-Hyun Lee b, Jeong-Su Kim b, Dong-Hun Shin c, Soo-Yong Lee b, Min Soo Cho d, Seung-Whan Lee d, Markus Reinthaler e, Jai-Wun Park e, Gi-Byung Nam d, Robert J Lederman f, Yonghyun Won g, June-Hong Kim b,
PMCID: PMC7591824  PMID: 33145462

Visual Abstract

graphic file with name fx1.jpg

Key Words: echocardiography, hypertrophic cardiomyopathy, mitral cerclage, radiofrequency ablation, ventricular septum

Abbreviations and Acronyms: AV, atrioventricular; CT, computed tomography; HCM, hypertrophic cardiomyopathy; RF, radiofrequency; TTE, transthoracic echocardiography

Highlights

  • Surgical myectomy is morbid and transcoronary alcohol septal ablation can result in geographic miss or occasional nontarget injury.

  • We developed a transvenous intraseptal radiofrequency ablation technique (cerclage ablation method) to overcome the shortcomings of surgical myectomy and alcohol septal ablation.

  • We delivered dedicated intraseptal radiofrequency ablation catheter into the ventricular septum through the coronary sinus and a septal vein. The ablated volume of myocardium was larger at higher electrode temperatures and using an irrigated-ablation mode in vitro. Cerclage ablation was successful in all in vivo attempts and induced a focal regional wall motion abnormality at later follow-up.

  • This study presents in vivo evidence of the feasibility, effectiveness, and safety of cerclage ablation method to debulk interventricular septal myocardium.

Summary

Debulking of left ventricular septal mass is typically accomplished using surgical myectomy, which is morbid, or using transcoronary alcohol septal ablation, which can result in geographic miss and occasional catastrophic nontarget coronary injury. The authors developed and tested operational parameters in vitro and vivo for a device to accomplish transvenous intraseptal radiofrequency ablation to reduce ventricular septal mass using a technique derived from mitral cerclage, which the authors call cerclage ablation. Cerclage ablation appeared feasible in vitro and safe and effective in vivo. Cerclage ablation is an attractive new approach to debulk the interventricular septum in obstructive hypertrophic cardiomyopathy. These data support clinical investigation.


Hypertrophic cardiomyopathy (HCM) causes heart failure, mitral valve regurgitation, and sudden cardiac death (1, 2, 3, 4). Disabling symptoms of left ventricular outflow tract obstruction can be relieved using surgical ventricular septal myectomy (5) and transcoronary alcohol septal ablation. Transcoronary alcohol septal ablation is an attractive alternative offered to older patients at higher risk of complications from surgical myectomy. Transcoronary alcohol septal ablation can result in geographic miss (when available septal perforator arteries fail to subtend the obstructing portion of the septum) and rare, but catastrophic, reflux (to the left anterior descending coronary artery) or spillover (for example, communicating to the posterior descending coronary artery), and more frequent requirement for permanent pacemaker implantation (6, 7, 8, 9, 10, 11, 12).

Recently Liu et al. (13,14) reported transthoracic needle radiofrequency (RF) ablation for HCM. Under general anesthesia and transthoracic echocardiography (TTE) guidance, a long 17G RF ablation needle was inserted into the hypertrophied interventricular septum through the chest wall and apical myocardium in 15 patients with HCM. Their approach risks epicardial coronary and myocardial injury, and requires traversing the chest wall and possibly lungs, along with general anesthesia. A transvenous catheter-based approach might be an attractive alternative that would obviate bystander tissue injury and general anesthesia. Our group has proposed a family of catheter procedures that enter the interventricular septum through the coronary veins to treat (mitral loop cerclage [15]), tricuspid valve regurgitation (cerclage tricuspid block [16]), and atrioventricular (AV) conduction failure (cerclage physiological pacing [17]). Coronary vein access also affords a novel opportunity to debulk the interventricular septum using RF ablation, which we describe here.

We developed a simple transvenous catheter approach to overcome the shortcomings of surgical myectomy and transcoronary alcohol septal ablation (Figure 1). We deliver a low-profile ablation catheter into the ventricular septum using the same transvenous coronary sinus technique and trajectory used in mitral cerclage annuloplasty (16,18). Herein we report the in vitro and in vivo evaluation of cerclage RF intraseptal ablation using purpose-built catheter prototypes.

Figure 1.

Figure 1

Cerclage RF Catheter System

(A) Multipolar type catheter design. (B) Monopolar type catheter design. (C) The prototype of the monopolar electrode catheter. (D) Overall picture of the catheter system. (E) Radiofrequency (RF) generator used in this study.

Methods

Prototype ablation devices

Monopolar (15-mm solenoid electrodes) or multipolar (5 to 6 electrodes spaced 3 mm apart) catheters were manufactured in open-lumen irrigated (5-Fr) and non-irrigated (4-F) forms, and were provided by Tau-PNU Medical (Busan, Republic of Korea) (7,19, 20, 21, 22). All incorporated a thermocouple immediately proximal to the electrodes. A dispersive electrode was applied externally to the skin. Irrigation used a commercially available saline pump (SP-8800, AMPall Co., Ltd, Seoul, Republic of Korea) operating 0.5 to 3ml/min. The ablation catheters were driven by a 400 kHz RF signal generator (Myogen V-1000, RF Medical, Ltd, Seoul, Republic of Korea) (Figure 1). The generator was operated in temperature-controlled mode and set to pause if impedance exceeded a threshold (≥250 ohms).

Ablation in vitro (ex vivo study)

To optimize ablation parameters, we tested the following variables: monopolar versus multipolar; nonirrigated versus irrigated (1 ml/min); ablation time (10, 20, or 30 min); and ablation temperature (65°C, 75°C, or 85°C). Experiments were performed in a constant-temperature 36°C saline-filled tank having a dispersion electrode at the bottom. On the basis of superior performance, temperature and duration experiments were confined to the monopolar irrigated configuration. We tested 6 sample swine heart explants for each of 9 parameters.

Animal experiments

Animal experiments were approved by the animal care and use committee policies of the Pusan National Yangsan University Hospital. General anesthesia was induced with alfaxalone 5 mg/kg and xylazine 2 mg/kg intramuscular, and maintained with isofluorane 3%. Amiodarone (150-mg bolus, then 1 mg/min) was administered intravenously as needed to suppress catheter-induced arrhythmia. A total of 11 (nonsurvival n = 4, and then survival n = 7) Yorkshire swine (45 ± 5.4 kg, Orient Bio, Seongnam, Republic of Korea) underwent cerclage RF ablation. The technique was developed and refined in the nonsurvival animals.

The cerclage RF intraseptal ablation procedure is based on the coronary sinus approach of mitral cerclage annuloplasty. After engaging the coronary sinus with a balloon-tipped guiding catheter (Cello catheter, 8-F, Medtronic, Dublin, Ireland), a pressurized radiocontrast venogram is performed to depict and select a target septal perforator vein branch. A commercial 0.014-inch percutaneous transluminal coronary angioplasty guidewire (Asahi Sion, Conquest wire, or Astato wire, all Asahi Intecc, Aichi, Japan) are used to engage the target vein with the help of a dual-lumen microcatheter (Crusade catheter, Kaneka Corporation, Tokyo, Japan). Once the wire is inside the myocardium through the septal vein, the wire is intentionally directed toward the target site, which, in most cases, requires the wire to pass through the wall of the septal vein. Traversing the myocardium with the guidewire is done easily with provocation of repetitious ventricular premature contractions. The wire is engaged deeply to provide catheter support, the position is confirmed with TTE, and then the cerclage RF catheter is advanced into position inside the interventricular septum over the wire. By virtue of the high-precision tapered tip together with the pushability-oriented catheter design of the braided shaft, countertraction of the guidewire in the right ventricular cavity was not needed in this study. We further tested the following parameters in vivo: ablation temperature, ablation duration, and irrigation versus nonirrigation.

Follow-up

Among survival animals, TTE was performed with a GE Vivid Q (GE Healthcare, Milwaukee, Wisconsin) using a 5 MHz transducer immediately and 2, 4, 6, and 8 weeks after the procedure. Parasternal long-axis and short-axis views were obtained and stored digitally for subsequent off-line analysis. All echocardiographic data were analyzed using an off-line analyzing software (EchoPAC PC v202, GE Healthcare) by an expert cardiologist (J.H. Park) in the core echocardiographic laboratory at the Chungnam National University Hospital). Measurement of myocardial thickness at the basal level was done at the end-diastolic frame more than 3 times, and an average value was used.

The primary endpoint of the animal experiments was technical feasibility of RF delivery into the intraseptal myocardium of the basal septum without major adverse events. Secondary endpoints of the survival animal experiments were safety and septal thinning during follow-up.

In the analysis of the wall-thinning data, >40% reduction was arbitrarily applied as the successful criteria of wall thinning on the basis of Liu’s human data (13). One animal underwent contrast-enhanced cardiac computed tomography (CT) 8 weeks after ablation. After euthanasia, cardiac explant specimens were formalin-fixed, paraffin-embedded, and then stained with hematoxylin and eosin for microscopic examination.

Statistical analysis

Continuous variables are presented as mean ± SD or median (25th to 75th percentiles) as determined by Kolmogorov-Smirnov normality test. Data were compared using Mann-Whitney U test, analysis of variance, or Welch Two Sample t-test as appropriate. Analyses were conducted using Statistical Package for the Social Sciences (IBM SPSS Statistics for Windows, version 21.0, released 2012, IBM Corp., Armonk, New York). A p value <0.05 was considered statistically significant.

Results

In vitro ablation parameters

In swine heart tissue treated in vitro, the impact of ablation parameters on tissue necrosis is summarized in Table 1. Monopolar and irrigated catheter prototypes caused larger ablation fields than multipolar and nonirrigated prototypes (550 ± 50 mm3 for monopolar vs. 290 ± 60 mm3 for multipolar, at 75°C for 10 min; p = 0.004). Ablation volume progressively increased after ablation durations increased from 10 to 20 min (820 ± 90 mm3 vs. 1,110 ± 150 mm3; p = 0.010, respectively), but not longer. Higher temperature and irrigation caused further incremental volume and diameter of ablation (1,430 ± 200 mm3 with irrigation vs. 280 ± 60 mm3 without irrigation, at a flow rate of 1 ml/min at 65°C; p < 0.001; 2,000 ± 210 mm3 with irrigation vs. 820 ± 90 mm3 without irrigation, at a flow rate of 1 ml/min at 75°C; p < 0.001) (Table 1, Figure 2). Regarding the diameter of the ablated tissue, the largest diameter was seen on temperature-controlled mode at 75°C with irrigation compared with the smallest diameter, which was seen on low temperature (65°C) controlled without irrigation using 5-F monopolar ablation for 10 min (13 ± 1 mm vs. 5 ± 0 mm; p < 0.001). The ablated diameter at 85°C without irrigation was between the value of the 75°C temperature control and 65°C under irrigation (12 ± 1 mm vs. 13 ± 1 mm vs. 10 ± 0 mm).

Table 1.

The Average of Ablated Myocardial Volume and Maximum Diameter of Volume According to Parameters in Ex Vivo Tests

Group Watts Average Resistance, Ω Ablated Myocardial Maximum Diameter of Volume, mm p Value Ablated Myocardial Volume, mm3 p Value
Type of catheter, 75°C, 10 min 4-F multipolar 1,571 93 5 ± 1 0.004 290 ± 60 0.004
4-F monopolar 2,185 68 8 ± 1 550 ± 50
Catheter size, 75°C, 10 min 4-F monopolar 2,185 68 8 ± 1 0.004 550 ± 50 0.004
5-F monopolar 2,343 63 9 ± 1 820 ± 90
Ablation time, 75°C, 5-F monopolar 10 min 2,343 63 9 ± 1 0.126 820 ± 90 0.006
20 min 4,060 70 10 ± 1 1,110 ± 150
30 min 5,512 63 10 ± 1 1,070 ± 180
Temperature, 10 min, 5-F monopolar 65°C 1,323 69 5 ± 0 <0.001 280 ± 60 <0.001
75°C 2,343 63 9 ± 1 280 ± 60
85°C 3,141 63 12 ± 1 1,620 ± 130
65°C
10 min, 5-F monopolar
No irrigation 1,323 69 5 ± 0 0.002§ 280 ± 60 <0.001§
Irrigation on, 1 ml/min 4,134 73 10 ± 0 1,430 ± 200
75°C
10 min, 5-F monopolar
No irrigation 2,343 63 10 ± 1 <0.001 820 ± 90
Irrigation on, 1 ml/min 4,798 72 13 ± 1 2,000 ± 210

Differences in volume or diameter according to temperature when performed with irrigation.

Differences in volume or diameter according to temperature when not performed with irrigation.

Difference in volume or diameter at 65°C depending on whether saline irrigation was used or not.

§

Difference in volume or diameter at 75°C depending on whether saline irrigation was used or not.

Difference between groups of 4.

Figure 2.

Figure 2

Ex Vivo Experiments to Determine Relationships Between RF Ablation and Myocardial Response

(A) Radiofrequency (RF) energy was delivered in temperature-controlled mode for 10 min with or without saline irrigation through the catheter (1 ml/min). (B) The result of ex vivo experiments to determine the relationships between RF ablation and swine myocardial response (p < 0.05 in differences among all groups of volume or maximal diameter of volume only except the difference between 75°C irrigation and 85°C nonirrigation groups). Red squares indicate the maximal diameter of the ablated myocardium.

Cerclage RF ablation in vivo

Efficacy

The cerclage RF ablation was successful in all swine (nonsurvival n = 4, survival n = 7). The average angiographic diameter of septal veins was 1.5 ± 0.4 mm, and the length was 6.6 ± 2.4 mm (Figure 3). Cerclage intraseptal RF application immediately induced akinesia in the target myocardium, which persisted through follow-up (Table 2). The 75°C or higher temperature-controlled mode with the 5-Fr irrigated monopolar catheter system showed more obvious wall thinning at 8 weeks’ follow-up.

Figure 3.

Figure 3

Representative Porcine Septal Venograms

The average diameter of the septal veins was 1.5 ± 0.4 mm, and the average length was 6.6 ± 2.4 mm (n = 10).

Table 2.

Echocardiographic Features in Survival Group After RF Ablation

Subject # Catheter Experimental Condition Watts Average Resistance, Ω Wall Thickness, mm
Wall Motion
QRS Width, ms
Baseline Edema After RF Follow-Up Immediate Post Follow-Up Baseline Follow-Up
1 4-F multipolar 15 min, 65°C
4 weeks follow-up
570 108 10.3 17.42 10.2 Thinning (+) NA 106
2 4-F multipolar 40 min, 65°C
8 weeks follow-up
2,396 73 11.5 15.77 7.6 Thinning (++) NA 75
3 4-F multipolar 60 min, 65°C
8 weeks follow-up
3,219 86 11.4 14.29 10.6 Thinning (+) +/- 67 72
4 4-F monopolar 30 min, 80°C
6 weeks’ follow-up
1,329 141 11.9 14.23 9.6 Thinning (+) 70 100
5 5-F monopolar 30 min, 85°C
8 weeks’ follow-up
2,259 65 11.6 20.02 7.4 Thinning (++) 70 107
6 5-F monopolar 45 min, 55°C
8 weeks’ follow-up
Performed with irrigation
8,114 87 12.5 13.37 11.4 Thinning (+) 92 82
7 5-F monopolar 15 min, 75°C
Sudden death at 2 weeks
Performed with irrigation
1,608 57 11.7 18.27 9.6 Thinning (+) 97 95

The wall motion of visual estimation in echocardiography subjectively determined, where − = no abnormality, −/+ = mild abnormality, Ⅰ = abnormality length was shorter than catheter length, Ⅱ = abnormality length same length of the catheter, and Ⅲ = abnormality length was longer than catheter length. Thinning (+) was defined as a reduction of a septal thickness, <30% in the baseline at last follow-up. Thinning (++) was defined as reduction of septal thickness by 30% or greater than baseline at last follow-up.

NA = not available due to technical problems; RF = radio frequency.

Two representative echocardiography-guided ablations in swine # 5 are shown in Figure 4. In all cases, ablation immediately induced a hyperechogenic tissue appearance and a regional wall motion abnormality that persisted at follow-up (Table 2, Figure 4). The wall motion score index by TTE decreased in 6 of the 7 swine in the survival group, after the procedure. Cardiac CT imaging also showed the ablation-induced septal thinning at 8 weeks’ follow-up.

Figure 4.

Figure 4

Serial Change of Septal Reduction After Cerclage RF Ablation

(A) Baseline (11.5 mm), (B) 2 weeks’ follow-up (9.4 mm), and (C) 8 weeks’ follow-up (7.6 mm, in subject #2). (D) Comparison at baseline and last follow-up of septal thickness in all animals (n = 7), (E) baseline (15.6 mm due to radiofrequency [RF]-induced edema, noninjured part 11.6 mm), (F) 8 weeks’ follow-up (7.4 mm), (G) long-axis view on computed tomography at 8 weeks (5.5 mm), and (H) short-axis view on computed tomography in subject #5.

Safety

During technical development experiments, ablation induced frequent ventricular premature contractions and/or ventricular fibrillation requiring defibrillation but were suppressed thereafter using parenteral amiodarone.

One animal (#7) died suddenly at 2 weeks, attributed to electrical instability from a large infarction. No other animals exhibited important physiological complications. Two animals (29%) exhibited transient AV block during RF ablation (n = 2). One animal (15%) exhibited QRS prolongation >20% (Table 2, Figure 5). No animal exhibited high-grade AV conduction block.

Figure 5.

Figure 5

ECG Data in the in Vivo Experiments

There was no high-degree atrioventricular block during and after the procedure, and there was a temporary slight QRS widening, but it recovered in all cases. The baseline electrocardiograms of cases #1 and #2 are missing data. ECG = electrocardiogram.

Histological findings

Histologic examinations were done at 8 weeks’ late follow-up (2 swine in survival), and a case of sudden death at 2 weeks (1 swine in survival). The specimen of late follow-up (survival #5 at 8 weeks) showed a large area of coagulated necrotic regions with borders that distinguished these regions from the surrounding normal healthy myocardial tissue (Figure 6). In subject #7, which showed sudden death at 2 weeks from excessive RF energy with respect to its baseline septal thickness, the pathological finding showed an acute ischemic injury pattern of the myocardium with consistent remote damage of the adjacent septal perforators up to 2.5 cm away from the direct ablation zone. Despite damage to the septal perforators, the epicardial left anterior descending artery was completely intact. In all specimens, tissue at the endocardial site appeared to be less damaged by RF injury.

Figure 6.

Figure 6

Gross Pathology and Histological Finding After Cerclage RF Septal Ablation

(A) The heart was harvested after 8 weeks’ follow-up and sliced at the target septum that showed a large area of coagulated necrotic region. The arrow indicates the ablated lesion. The boxed area is shown enlarged in B. (B) The asterisk indicates the site of the ablation catheter. The boxed area is shown in greater magnification in C. Judging from the catheter position, asymmetric ablation has occurred in favor of endocardial sparing with regard to the ablated territory. (C) The heart sliced at the target septum showed a large area of coagulated necrotic region. (D) The left anterior descending coronary artery (arrow) close to the ablated zone was patent. (E) In a beating heart experiment, using an ablation catheter with a 15-mm–long electrode coil, when ablation was done for 15 min at 75°C with saline irrigation through the catheter (1 ml/min), the length of the ablated region was about 15 mm and the width was about 12 mm. RF = radiofrequency.

Discussion

This study shows that targeted cerclage intraseptal RF ablation accomplishes septal debulking via a transvenous approach, without inducing high-grade AV block. Endocardial RF ablation has been described previously for HCM treatment (23, 24, 25, 26, 27, 28), and it usually, if not universally, induces high-grade AV conduction block requiring permanent pacing.

Other approaches to RF ablation have important shortcomings. Endocardial ablation using 7- to 8-F right- or left-ventricular cardiac electrophysiology catheters requires multiple points of ablation contact that may be discontinuous, may ablate excessive nontarget myocardium, and confers a high risk of high-grade AV conduction block (29). Recently, Liu et al. (13) reported transapical needle intraseptal RF ablation, which is unattractive in requiring protracted transthoracic transapical needle access to the basal septum despite cardiac and respiratory motion.

Our transvenous cerclage intraseptal RF ablation approach has other potential advantages. Because it is “right-sided,” it does not risk thromboembolic complications including stroke. And guidewires are navigated through coronary veins to intended targets. This characteristic of flexible manipulation of wire navigation to the intended target cannot deny the benefit of suitable anatomy of the septal perforator vein for an easier procedure compared with the case of having a poor septal vein. The target territory can be confirmed when the guidewire or catheter is visualized via echocardiography before ablation.

Prototype of ablation catheter and methods of RF delivery

Our prototype 4- to 5-F cerclage ablation catheters were easily advanced to intraventricular septal targets without using accessories beyond guiding sheaths, catheters, and guidewires, analogous to techniques used to introduce cardiac resynchronization therapy electrodes.

With this system (Table 1), we are able to ablate a large volume of myocardium (up to 1,624.1 ± 131.1 mm3) along the axis of the ablation catheter. As expected, ablation performance is superior using monopolar prototypes, with irrigation, and with longer and higher temperature settings. We predict this performance will allow better access to the basal septum compared with current (endocardial and transcoronary alcohol septal) ablation techniques.

As for imaging guidance for this procedure, a pre-procedural cardiac CT scan usually can show the anatomy of the septal vein very well and should be an important imaging tool to determine and plan the procedure beforehand (15). The whole procedure can be done under fluoroscopic guidance only. However, TTE or transesophageal echocardiography guidance is very useful to check the guidewire position during the procedure, which is important in order to have a successful and safe procedure.

Potential application of cerclage RF ablation

The novel approach of our study can be applied, not only to HCM treatment, but also to other diseases that mandate septal reduction or ablation (8,17). For instance, obstructive sigmoid septum without hypertrophied septum for focal RF ablation, ventricular tachycardia ablation in the basal interventricular septum, or preemptive septal reduction for relieving left ventricular outflow tract obstruction risk in transcatheter mitral valve replacement could be considered as potential indications of this approach.

Study limitations

First, because we do not have an animal model of the hypertrophied septum, the normal septum was tested in our study. However, hypertrophied septum might have more room for placement of the ablation catheter and theoretically would also be more tolerant to the risk of conduction system damage. Second, this study was done only for proof of our concept for procedural feasibility with a handful of animal experiments; the actual response and long-term follow-up could be different in real-world clinical use in patients. Dosimetry data from the ex vivo study for looking at the relationship between ablation parameters and amount ablation tissue may not match with the in vivo study. It should be kept in mind that the amount of ablated tissue achieved in perfused tissue in a living organ is known to be approximately 30% lower than that of nonperfused ex vivo tissue (30). Devices used in this study are not the final mature model for human translation.

Conclusions

The findings confirm the feasibility, effectiveness, and safety of transvenous ventricular septal RF ablation along a mitral cerclage trajectory to debulk interventricular septal tissue. This promising preclinical result supports further investigation to debulk or prevent left ventricular outflow tract obstruction such as in hypertrophic cardiomyopathy or transcatheter mitral valve implantation.

Perspectives.

COMPETENCY IN MEDICAL KNOWLEDGE 1: Transcoronary alcohol septal ablation can result in geographic miss and occasional nontarget injury. Surgical myectomy is morbid. Alternative septal ablation strategies are desirable. Transvenous radiofrequency ablation of the septum, using the mitral cerclage technique, can accomplish ventricular septal ablation without geographic miss and without risking nontarget injury.

TRANSLATIONAL OUTLOOK: A future study needs to confirm these promising results in the clinical praxis.

Author Relationship With Industry

This study was supported by Research and Development grant S2773417 from the Korean Ministry of Small and Medium Enterprise and Startups. Tau-PNU Medical provided the prototype ablation catheters. Dr. J.-W. Park is a medical consultant for Tau-PNU Medical. Drs. J.-H. Kim and Nam are coinventors on patents for devices for mitral loop cerclage ablation. Dr. J.-H Kim is a stockholder of Tau-PNU Medical. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.

Acknowledgments

The authors are grateful to Jun-Oh Kim for procedure assistance, and to Su-Jin Jung and So Hee Nam for data analysis.

Footnotes

The authors attest they are in compliance with human studies committees and animal welfare regulations of the authors’ institutions and Food and Drug Administration guidelines, including patient consent where appropriate. For more information, visit the JACC: Basic to Translational Scienceauthor instructions page.

References

  • 1.Elliott P., Gimeno J., Tomé M., McKenna W. Left ventricular outflow tract obstruction and sudden death in hypertrophic cardiomyopathy. Eur Heart J. 2006;27:3073. doi: 10.1093/eurheartj/ehl383. [DOI] [PubMed] [Google Scholar]
  • 2.Maron M.S., Olivotto I., Betocchi S. Effect of left ventricular outflow tract obstruction on clinical outcome in hypertrophic cardiomyopathy. N Engl J Med. 2003;348:295–303. doi: 10.1056/NEJMoa021332. [DOI] [PubMed] [Google Scholar]
  • 3.Gimeno J.R., Tome-Esteban M., Lofiego C. Exercise-induced ventricular arrhythmias and risk of sudden cardiac death in patients with hypertrophic cardiomyopathy. Eur Heart J. 2009;30:2599–2605. doi: 10.1093/eurheartj/ehp327. [DOI] [PubMed] [Google Scholar]
  • 4.Elliott P.M., Anastasakis A., Borger M.A. 2014 ESC guidelines on diagnosis and management of hypertrophic cardiomyopathy: the Task Force for the Diagnosis and Management of Hypertrophic Cardiomyopathy of the European Society of Cardiology (ESC) Eur Heart J. 2014;35:2733–2779. doi: 10.1093/eurheartj/ehu284. [DOI] [PubMed] [Google Scholar]
  • 5.Armistead S., Williams B. Hypertrophic obstructive cardiomyopathy. The use of a diathermy loop for septal resection. J Cardiovasc Surg. 1984;25:185–186. [PubMed] [Google Scholar]
  • 6.Smedira N.G., Lytle B.W., Lever H.M. Current effectiveness and risks of isolated septal myectomy for hypertrophic obstructive cardiomyopathy. Ann Thorac Surg. 2008;85:127–133. doi: 10.1016/j.athoracsur.2007.07.063. [DOI] [PubMed] [Google Scholar]
  • 7.Iacovoni A., Spirito P., Simon C. A contemporary European experience with surgical septal myectomy in hypertrophic cardiomyopathy. Eur Heart J. 2012;33:2080–2087. doi: 10.1093/eurheartj/ehs064. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Kim L.K., Swaminathan R.V., Looser P. Hospital volume outcomes after septal myectomy and alcohol septal ablation for treatment of obstructive hypertrophic cardiomyopathy: US Nationwide Inpatient Database, 2003-2011. JAMA Cardiol. 2016;1:324–332. doi: 10.1001/jamacardio.2016.0252. [DOI] [PubMed] [Google Scholar]
  • 9.Chan W., Williams L., Kotowycz M.A. Angiographic and echocardiographic correlates of suitable septal perforators for alcohol septal ablation in hypertrophic obstructive cardiomyopathy. Can J Cardiol. 2014;30:912–919. doi: 10.1016/j.cjca.2014.04.008. [DOI] [PubMed] [Google Scholar]
  • 10.Alam M., Dokainish H., Lakkis N. Alcohol septal ablation for hypertrophic obstructive cardiomyopathy: a systematic review of published studies. J Interv Cardiol. 2006;19:319–327. doi: 10.1111/j.1540-8183.2006.00153.x. [DOI] [PubMed] [Google Scholar]
  • 11.Liebregts M., Faber L., Jensen M.K. Outcomes of alcohol septal ablation in younger patients with obstructive hypertrophic cardiomyopathy. J Am Coll Cardiol Intv. 2017;10:1134–1143. [Google Scholar]
  • 12.Liebregts M., Vriesendorp P.A., Ten Berg J.M. Alcohol septal ablation for obstructive hypertrophic cardiomyopathy: a word of endorsement. J Am Coll Cardiol. 2017;70:481–488. doi: 10.1016/j.jacc.2017.02.080. [DOI] [PubMed] [Google Scholar]
  • 13.Liu L., Li J., Zuo L. Percutaneous intramyocardial septal radiofrequency ablation for hypertrophic obstructive cardiomyopathy. J Am Coll Cardiol. 2018;72:1898–1909. doi: 10.1016/j.jacc.2018.07.080. [DOI] [PubMed] [Google Scholar]
  • 14.He G., Sun C., Zhang X. Echocardiography-guided percutaneous per-ventricular laser ablation of ventricular septum: in vivo study in a canine model. Lasers Med Sci. 2016;31:645–651. doi: 10.1007/s10103-016-1881-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Kim J.H., Kocaturk O., Ozturk C. Mitral cerclage annuloplasty, a novel transcatheter treatment for secondary mitral valve regurgitation: initial results in swine. J Am Coll Cardiol. 2009;54:638–651. doi: 10.1016/j.jacc.2009.03.071. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Chon M.-K., Jung S.-M., Lee S.Y. TCT-18 Novel concept of catheter-based treatment for tricuspid regurgitation (Cerclage-TR block): a preliminary animal experiment in a swine model (abstr) J Am Coll Cardiol. 2018;72 Suppl:B8. [Google Scholar]
  • 17.Cho M.S., Chon M.K., Choi J.H. Cerclage parahisian septal pacing through the septal perforator branch of the great cardiac vein: bedside-to-bench development of a novel technique and lead. J Heart Rhythm. 2019;16:1834–1840. doi: 10.1016/j.hrthm.2019.08.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Park Y.-H., Chon M.-K., Lederman R.J. Mitral loop cerclage annuloplasty for secondary mitral regurgitation: first human results. J Am Coll Cardiol Intv. 2017;10:597–610. doi: 10.1016/j.jcin.2016.12.282. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Mulier P.M.J., Hoey M.F., inventors; DUTHLER REED A, Medtronic Inc, assignee . US patent 5431649A; 1997. Method and apparatus for R-F ablation. [Google Scholar]
  • 20.Rydell M.A., inventor; Everest Medical Corp, assignee . CA patent 2037242A1; 1992. RF ablation catheter. [Google Scholar]
  • 21.Imran M.A., Pomeranz M.L., inventors; Boston Scientific Corp, Boston Scientific Scimed, assignees . US patent 5348554A; 1994. Catheter for RF ablation with cooled electrode. [Google Scholar]
  • 22.Müssigbrodt A., Grothoff M., Dinov B. Irrigated tip catheters for radiofrequency ablation in ventricular tachycardia. Biomed Res Int. 2015;2015:389294. doi: 10.1155/2015/389294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Cooper R.M., Shahzad A., Hasleton J. Radiofrequency ablation of the interventricular septum to treat outflow tract gradients in hypertrophic obstructive cardiomyopathy: a novel use of CARTOSound® technology to guide ablation. Ep Europace. 2016;18:113–120. doi: 10.1093/europace/euv302. [DOI] [PubMed] [Google Scholar]
  • 24.Crossen K., Jones M., Erikson C. Radiofrequency septal reduction in symptomatic hypertrophic obstructive cardiomyopathy. Heart Rhythm. 2016;13:1885–1890. doi: 10.1016/j.hrthm.2016.04.018. [DOI] [PubMed] [Google Scholar]
  • 25.Lawrenz T., Borchert B., Leuner C. Endocardial radiofrequency ablation for hypertrophic obstructive cardiomyopathy: acute results and 6 months' follow-up in 19 patients. J Am Coll Cardiol. 2011;57:572–576. doi: 10.1016/j.jacc.2010.07.055. [DOI] [PubMed] [Google Scholar]
  • 26.Poon S.S., Cooper R.M., Gupta D. Endocardial radiofrequency septal ablation—a new option for non-surgical septal reduction in patients with hypertrophic obstructive cardiomyopathy (HOCM)?: a systematic review of clinical studies. Int J Cardiol. 2016;222:772–774. doi: 10.1016/j.ijcard.2016.08.123. [DOI] [PubMed] [Google Scholar]
  • 27.Sreeram N., Emmel M., de Giovanni J.V. Percutaneous radiofrequency septal reduction for hypertrophic obstructive cardiomyopathy in children. J Am Coll Cardiol. 2011;58:2501–2510. doi: 10.1016/j.jacc.2011.09.020. [DOI] [PubMed] [Google Scholar]
  • 28.Yang H., Yang Y., Xue Y., Luo S. Efficacy and safety of radiofrequency ablation for hypertrophic obstructive cardiomyopathy: a systematic review and meta-analysis. Clinical Cardiology. 2020;43:450–458. doi: 10.1002/clc.23341. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Kaye G.C., Linker N.J., Marwick T.H. Effect of right ventricular pacing lead site on left ventricular function in patients with high-grade atrioventricular block: results of the Protect-Pace study. Eur Heart J. 2015;36:856–862. doi: 10.1093/eurheartj/ehu304. [DOI] [PubMed] [Google Scholar]
  • 30.Ong S.L., Gravante G., Metcalfe M.S., Dennison A.R. History, ethics, advantages and limitations of experimental models for hepatic ablation. World J Gastroenterol. 2013;19:147–154. doi: 10.3748/wjg.v19.i2.147. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from JACC: Basic to Translational Science are provided here courtesy of Elsevier

RESOURCES