Table 2.
Time of consecutive applications of defibrillations.
| No.∗ | Group MFE | Group C | p value |
|---|---|---|---|
| 1st defibrillation | 39.0 ± 10.6 (100)∗∗ | 50.4 ± 16.8 (100) | <0.001 |
| 2nd defibrillation | 166.6 ± 23.0 (100) | 179.8 ± 28.1 (100) | 0.011 |
| 3rd defibrillation | 290.2 ± 24.0 (100) | 313.1 ± 39.6 (100) | 0.001 |
| 4th defibrillation | 413.3 ± 25.1 (100) | 440.9 ± 49.1 (100) | 0.001 |
| 5th defibrillation | 535.9 ± 24.5 (100) | 535.1 ± 63.0 (86) | 0.785 |
∗∗Data are expressed as mean with standard deviation (sd); ∗number of analysed defibrillations. C = control; MFE = multifunction electrode.