Skip to main content
. 2020 Aug 12;123(9):1437–1444. doi: 10.1038/s41416-020-1024-2

Table 1.

Analysis accuracy of EGFR L858R, 19Del and T790M in dEGFR39 assay was determined by testing the DNA Reference Standards.

19Del L858R T790M
Expected mutant ratio 12.50% 12.50% 12.50%
Measured mean ± SD 12.63 ± 1.05% 12.13 ± 0.97% 12.19 ± 1.03%
CV 0.08 0.08 0.08
Paired t-Test
 P-value (vs. Expected) 0.70 0.24 0.36
 Significantly different (P < 0.05) NO NO NO
Bland–Altman analysis
 Bias ± SD 0.13 ± 1.05% 0.37 ± 0.97% 0.31 ± 1.03%
 95% CI −0.57 to 0.83% −1.53 to 2.27% −1.71 to 2.33%

SD standard deviation, CV coeffiency variation, CI confidence interval.