Table 1. .
The preset cone-beam CT (CBCT) scan protocols.
| Low dose head | Standard dose head | Low dose thorax | Pelvis | |
| X-ray voltage (kVp) | 100 | 100 | 110 | 125 |
| X-ray current (mA) | 10 | 20 | 20 | 80 |
| X-ray millisecond (ms) | 20 | 20 | 20 | 13 |
| Gantry rotation (deg) | 200 | 200 | 360 | 360 |
| Number of projections | 360 | 360 | 655 | 655 |
| Exposure (mAs) | 72 | 145 | 262 | 680 |
| Fan type | Full fan | Full fan | Half fan | Half fan |
| Bow-tie filter | Full | Full | Half | Half |
| Default pixel matrix | 384 × 384 | 384 × 384 | 384 × 384 | 384 × 384 |
| Dose CTDIw (mGy) | 2 | 3.9 | 4.7 | 17.7 |
| Recon. filter | Standard | Sharp | Standard | Standard |
| Slice thickness (mm) | 2.5 | 2.5 | 2.5 | 2.5 |