Skip to main content
. 2020 Oct 15;11:576729. doi: 10.3389/fneur.2020.576729

Figure 10.

Figure 10

(A) Different types of filters that can be used to shape the EMG spectrum, to keep, remove, or attenuate certain frequency components of the EMG signal + noise. In the example shown, the low-pass filter has a cut-off frequency (fc) of 170 Hz and the high-pass filter has a cut-off frequency of 140 Hz. The band-pass and notch filters have lower cut-off frequencies (fc1) of 70 Hz and upper cut-off frequencies (fc2) of 108 Hz. (B) Schematic of the typical stages in recording surface EMG signals, with filtering at several points along the process (filters that operate on the analog signal, i.e., hardware filters). With active electrodes, pre-amplification is performed within the electrode itself (rather than the amplification being performed in an external circuit), see section Choice of Amplifier. EMG signals are low-pass filtered before sampling to suppress high-frequency components and prevent the distortion of the spectral content, see Figure 8. See Examples (ix) and (x) in Tutorial Code. (C) An example of a noisy sEMG signal contaminated with 50 Hz interference, the frequency response of a 50 Hz notch filter and a 20–500 Hz band-pass filter (the figure indicates how much the sEMG signal is attenuated by the filter, in dB, at each frequency), and filtered sEMG spectrum (after notch and band-pass filtering the raw sEMG signal). Note that notch filters are typically used when only an approximate estimate of EMG amplitude is required.