Negative Sample OD |
95% prediction interval of 0.000 to 0.250 |
Limit of Detection |
52.85 ELISA units/mL |
Lower Limit of Quantification |
54.87 ELISA units/mL |
Upper Limit of Quantification |
30,632.33 ELISA units/mL |
Residual (Repeatability) |
15.3% CV |
Sum of Operator, Day Plate, Replicate Effects (Intermediate Precision) |
19.9% CV |
Total Assay Variability Due to Method (Intermediate Precision and Repeatability) |
25.3% CV |
Dilutional Linearity (Accuracy) |
Slope of -0.99 with 90% confidence of (-1.02, -0.97) |
Selectivity (Matrix Effects) |
Percent difference (-5.1%, 13.2%). |
Specificity |
Mean ratio of Mock to 25 μg/mL rGP = 5.05 |
Mean ratio of Mock to 10 μg/mL rGP = 4.06 |
Mean ratio of Mock to 25 μg/mL CMV = 1.06 |